Batteries

X Series models use an easily replaced rechargeable lithium-ion battery pack (the SurePower II
Battery Pack). A new, fully charged battery pack typically delivers more than 6 hours of ECG
monitoring. Use of other functions (such as the defibrillator, printer, or pacemaker) reduces this
time.
When a LOW BATTERY icon appears on the display and the unit emits three beeps in
conjunction with the displayed battery icon, the battery must be replaced and recharged.
You can charge the battery by either of the following methods:
Internal charging — plug the X Series into an auxiliary power adapter to automatically
begin charging the installed battery pack. The front panel battery indicator operates as
follows:
Note:
External charging — use the ZOLL SurePower Battery Charger with the X Series battery
adapter to charge the battery pack and test the battery's capacity. For details, refer to the
SurePower II Battery Pack Guide.
The Recalibration LED icon ( ) lights for approximately 10 seconds (after you press and
release the Display button) if the battery needs to be calibrated. If the Recalibration LED lights,
the runtime indicator will not display run time for that battery. For best performance of the
battery, you should recalibrate the battery as soon as possible.
To manually recalibrate the SurePower II Battery Pack, you can insert the battery into the
SurePower Charger Station and perform a Manual Test (for more information, see the ZOLL
SurePower Charger Station Operator's Guide).
After you recalibrate the battery, the Recalibration LED will only flash when you press the
Display button.
9650-001355-01 Rev. M
When the indicator is:
Steady yellow
Steady green
Alternating yellow and
green
Not lit
Upon power up, it takes approximately 45 seconds for the LEDs on the battery to
accurately display run time.
X Series Operator's Guide
X Series Product Functions
It means:
Battery is charging.
Battery is charged.
The charge state cannot be
determined or a battery charging
fault has been detected.
No battery in device.