Press Analyze Button - ZOLL X Series Operator's Manual

Portable defibrillator, includes real cpr help and see-thru cpr
Table of Contents

Advertisement

Chapter 17 Analysis/CPR Protocol Defibrillation

2 Press ANALYZE Button

WARNING!
Keep patient motionless during ECG analysis. Do not touch the patient during analysis.
Cease all movement via stretcher or vehicle before analyzing the ECG.
Press the ANALYZE button to begin the analysis of the patient's ECG rhythm and to determine
if a shockable rhythm is present.
The ANALYZING ECG and STAND CLEAR messages alternate at the bottom of the screen
while the patient's ECG is analyzed. Once the analysis is completed, the unit indicates whether
or not a shock is advised.
Note:
The analysis determines whether there is a shockable rhythm. If the analysis detects a
shockable rhythm the unit prompts the operator to shock the patient at the preconfigured energy
level. If the analysis does not detect a shockable rhythm, the unit alerts the operator that no
shock is advised.
WARNING!
ECG rhythm analysis does not warn of patient asystole, which is not a shockable
rhythm.
17-4
You can terminate the Analysis/CPR Protocol mode and return to Manual mode at any
time by pressing the Exit quick access key.
www.zoll.com
9650-001355-01 Rev. M