Spo2 Monitor Test - ZOLL M Series Service Manual

Hide thumbs Also See for M Series:
Table of Contents

Advertisement

M Series Service Manual
17.0 SpO
Monitor Test
2
Tools Needed:

Masimo
Reusable Sensor.
Masimo
Patient Cable
Fluke Index Series 2 SpO
Simulator.
2
Test Setup:
Connect the universal cable to the MFC test plug.
DO NOT connect the ECG cable to the simulator.
Connect the Masimo
Patient Cable and attach the Masimo
Connect the Masimo
sensor to the finger simulation post.
Place a fully charged battery into the battery well or connect to ac power (dc power, if equipped).
Ensure that the SpO
Simulator is off.
2
Do this...
17.1
Turn the selector switch to MONITOR.
(For AED units, turn the selector switch to ON
and select Manual mode.)
17.2
Wait ten seconds.
Turn on the SpO 2 simulator. Press the SIM
softkey on the Index SpO
the MAN softkey.
17.3
Press the 02+ or 02- softkey of the simulator
until the SpO
output is at 98%.
2
1-32
sensor to the patient cable.
Verify that...
The SpO
The unit displays the SpO
Simulator, then press
2
The SpO
Note: You may need to wait up to 2 minutes for the information to appear on
the ZOLL display.
saturation percentage appears as a dashed line on the monitor.
2
PULSE SEARCH message.
2
reading of 98 +/- 1% appears on the M Series monitor.
2
Pass/Fail
o o
o o
o o
9650-0450-01 Rev. U

Hide quick links:

Advertisement

Table of Contents
loading