Specifications
Respiration Range: 0 to 149 breaths per minute
Respiration Rate Accuracy:
0 to 70 bpm:
71 to 120 bpm: ±2 bpm
121 to 149 bpm: ±3 bpm
Flow rate: 50 ml/min -7.5 + 15 ml/min, flow measured by volume.
Total System Response Time: 2.9 seconds typical, 3.9 seconds maximum.

Pulse Oximeter

Range:
Accuracy:
A-20
±1 bpm
Oxygen Saturation (% SpO
Carboxyhemoglobin Saturation (% SpCO)
Methemoglobin Saturation (% SpMet)
Total Hemoglobin (g/dL SpHb)
Total Oxygen Content (% SpOC)
Perfusion Index (% PI)
Pleth Variability Index (% PVI)
Pulse Rate (bpm)
Oxygen Saturation (% SpO
60 -80 ± 3%, Adults/Pediatrics/Infants
70 - 100 ± 2%, Adults/ Pediatrics/Infants; ± 3%, Neonates
Oxygen Saturation (% SpO
70% - 100% ±3% Adults/ Pediatrics/Infants/Neonates
Oxygen Saturation (% SpO
70 - 100 ± 2%, Adults/Pediatrics/Infants/Neonates
Pulse Rate (bpm) - During No Motion Conditions
25 - 240 ±3 bpm Adults/Pediatrics/Infants/Neonates
Pulse Rate (bpm) - During Motion Conditions
25 - 240 ±5 bpm Adults/Pediatrics/Infants/Neonates
Pulse Rate (bpm) - During Low Perfusion Conditions
25 - 240 ±3 bpm Adults/Pediatrics/Infants/Neonates
Carboxyhemoglobin Saturation (% SpCO)
1% - 40% ±3% Adults/Pediatrics/Infants
Methemoglobin Saturation (% SpMet)
1% - 15% ±1% Adults/Pediatrics/Infants/Neonates
Total Hemoglobin (ml/dL SpHb)
8 – 17 ±1 g/dL (arterial or venous) Adults/Pediatrics
Resolution:
SpO
: 1%
2
SpCO: 1%
SpMet: 0.1% for range up to 9.9%, 1% for range 10 - 99%
SpHb: 0.1 g/dL
SpOC: 0.1 ml/dL
PVI: 1%
PI: 0.1%
Pulse rate: 1 bpm (beats per minute)
www.zoll.com
)
0% - 100%
2
0% - 99%
0% - 99%
0 – 25 g/dL
0 – 35 ml/dL
0.02% - 20%
0% - 100%
25 - 240 beats per minute
) - During No Motion Conditions
2
) - During Motion Conditions
2
) - During Low Perfusion Conditions
2
9650-001355-01 Rev. M