Administrator's Guide
9650-000752-01 Rev. D

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the AED 3 and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for ZOLL AED 3

  • Page 1 Administrator’s Guide 9650-000752-01 Rev. D...
  • Page 2 The issue date for the ZOLL AED 3 Administrator’s Guide (REF 9650-000752-01 Rev. D) is November, 2016. Copyright © 2016 ZOLL Medical Corporation. All rights reserved. AED Pro, CPR Uni-padz, CPR-D-padz, Pedi-padz, Real CPR Help, Rectilinear Biphasic, RescueNet, Stat-padz, ZOLL, and ZOLL AED 3 are trademarks or registered trademarks of ZOLL Medical Corporation in the United States and/or other countries.
  • Page 3: Table Of Contents

    Table of Contents Preface Conventions..................iv ZOLL AED 3 Safety Considerations ............. iv Tracking Requirements ..............iv Notification of Adverse Events ............v Technical Service ................v International Customers ............vi Software License Agreements............vi Chapter 1 Overview ZOLL AED 3 User Interface..............2 Child Mode..................
  • Page 4 Cleaning the ZOLL AED 3 Defibrillator ..........24 Optional Maintenance Test for Technical Professionals...... 25 Test Procedure (ZOLL AED 3 and ZOLL AED 3 BLS Models) ... 25 Test Procedure (ZOLL AED 3 Automatic Models) ......26 Battery Maintenance ............... 27 Identifying Battery Conditions ............
  • Page 5: Preface

    AHA/ERC/ILCOR Guidelines for use of AEDs. This guide provides information about the operation and care of the ZOLL AED 3 defibrillator. The ZOLL AED 3 Administrator’s Guide is to be used by trained medical personnel in conjunction with the ZOLL AED 3 Operator’s Manual.
  • Page 6: Conventions

    Appendix C—Contains definitions of all configuration settings for the ZOLL AED 3 defibrillator and provides instructions for setting up your wireless configuration. Conventions Throughout this document, text prompts and voice prompts are indicated by capital italicized letters, such as CALL FOR HELP.
  • Page 7: Notification Of Adverse Events

    • Description of the problem • If you need to send the ZOLL AED 3 defibrillator to ZOLL Medical Corporation, obtain a service order request number from the Technical Service representative. NOTE Before sending the defibrillator and battery to ZOLL, check with your local shipping company to see if there are any restrictions on shipping lithium batteries.
  • Page 8: International Customers

    AED and return the AED and battery pack in its original container (or equivalent packaging) to the nearest authorized ZOLL Medical Corporation Service Center. To locate an authorized service center, contact the nearest ZOLL sales office or authorized distributor. For a list of ZOLL offices, please visit: http://www.zoll.com/contact/worldwide-locations/.
  • Page 9: Chapter 1 Overview

    Real CPR Help for CPR monitoring and feedback. • ZOLL AED 3 BLS — Operates in Semi-Automatic Mode. Includes Real CPR Help for CPR monitoring and feedback, optional patient ECG display, CPR Dashboard, optional lay rescuer graphics, and optional audio recording.
  • Page 10: Zoll Aed 3 User Interface

    The steps that include graphics and audio prompts are defined by local protocols for use of an AED. Child Mode The ZOLL AED 3 defibrillator has a Child mode that can be enabled in two ways: ® when the rescuer uses ZOLL CPR Uni-padz defibrillation pads and presses the ®...
  • Page 11: Using Defibrillation Pads

    If the defibrillation pads cable is not attached properly, the AED issues the PLUG IN PADS CABLE prompt. Make certain to connect the defibrillation pad cable to the ZOLL AED 3 defibrillator and attach the pads to the patient properly. WARNING! Defibrillation pads must be pre-attached to the device.
  • Page 12: Data Recording And Storage

    Transferring Files Using a USB Flash Drive All ZOLL AED 3 models have a USB port that allows you to export and save configuration settings, clinical archives, and device history to a USB flash drive. This feature allows you to easily copy a configuration from one AED to other AEDs.
  • Page 13: Zoll Aed 3 Operation

    Defibrillation Pad Connector — This connector is used to attach defibrillation pads to the AED. 9650-000752-01 Rev. D ZOLL AED 3 Administrator’s Guide...
  • Page 14: Graphical Display Area

    Support Bar — The brace that opens to hold the AED in an upright position. Graphical Display Area The ZOLL AED 3 defibrillator has a 5.3 x 9.5 centimeter LCD screen that displays the following information in the modes in this section: AED Management Mode and Rescue Mode.
  • Page 15: Rescue Mode

    Rescue Mode The ZOLL AED 3 and ZOLL AED 3 Automatic models show the Lay Rescuer display, while the ZOLL AED 3 BLS model can be configured to show the Lay Rescuer display, CPR only display, or CPR and ECG display.
  • Page 16 CPR Only/CPR and ECG Displays (ZOLL AED 3 BLS models only) Chest compression rate CPR countdown timer Rate Depth CPR Time (in) (cpm) 0:15 Chest compression ECG rhythm depth (optional) ECG scale marker User Number of prompt GOOD COMPRESSIONS shocks...
  • Page 17: Audio Prompts

    Audio Prompts During clinical use of the ZOLL AED 3 defibrillator, you may hear the following audio prompts: Table 1. Clinical Audio Prompts Audio Prompt Definition/Action AUTOMATIC DEFIBRILLATOR The AED is the fully automatic model that has the capability to automatically shock the patient without the rescuer having to press the Shock button.
  • Page 18 Release the Shock Button and press again after the ready tone sounds. SHOCK WILL BE DELIVERED (ZOLL AED 3 Automatic model only) The AED will IN THREE, TWO, ONE automatically deliver shock after the countdown completes.
  • Page 19 Stop CPR. The AED is about to begin an ECG rhythm analysis. CPR ONLY MODE Defibrillation is not working properly; AED will only provide CPR support. Prompts that you may hear during non-clinical use of the ZOLL AED 3 defibrillator include: Table 2. Non-Clinical Audio Prompts Audio Prompt Definition...
  • Page 20 ZOLL AED 3 Administrator’s Guide 9650-000752-01 Rev. D...
  • Page 21: Chapter 2 Data Transfer

    USB flash drive. This allows you to quickly and easily update software or copy a configuration to multiple AEDs. In order to transfer data, the ZOLL AED 3 defibrillator must be in AED Management mode. To enter AED Management Mode, press and hold the On/Off button for at least 5 seconds, then press the Device Configuration icon on the main display screen.
  • Page 22: Data Storage

    The ZOLL AED 3 BLS model has an audio recording option that can be configured on or off. If the audio recording option is on, the ZOLL AED 3 BLS model can record and store 60 minutes of continuous audio, ECG, and clinical event data.
  • Page 23: Exporting Files From The Aed

    Press OK to return to the previous screen. NOTE If you rename a configuration file with the intention of importing it into another ZOLL AED 3 defibrillator, do not include spaces in the file name. 9650-00752-01 Rev. D ZOLL AED 3 Administrator’s Guide...
  • Page 24: Using Rescuenet Code Review Software

    ZOLL Online. Using its Wi-Fi capability, the ZOLL AED 3 can be configured to transmit clinical event data over Wi-Fi to ZOLL Case Review. When using Case Review software to manage clinical event data, you can perform the following functions: •...
  • Page 25: Importing Files

    Press the Import Files icon and select one the following options to import: Configuration • • System Software Import Files • Root Certificates icon A green check displays next to the option you have selected. 9650-00752-01 Rev. D ZOLL AED 3 Administrator’s Guide...
  • Page 26: Troubleshooting Communication Issues

    Troubleshooting Communication Issues The following table summarizes communication messages that the ZOLL AED 3 defibrillator may issue, and their associated description and/or recommended action. If the AED is not working properly, contact ZOLL’s Technical Service Department for assistance. See “Technical Service” on page v for more information.
  • Page 27 THAT THE USB FLASH DRIVE IS PROPERLY INSERTED. UNABLE TO TRANSMIT The attempt to export the file to the USB REQUESTED FILE. flash drive was unsuccessful. Try the file transfer again. 9650-00752-01 Rev. D ZOLL AED 3 Administrator’s Guide...
  • Page 28 SYSTEM WILL RESTART. THE restarts. SCREEN WILL GO BLANK DURING THIS PROCESS. THIS MAY TAKE UP TO 5 MINUTES. DO NOT POWER OFF DURING THIS TIME. ZOLL AED 3 Administrator’s Guide 9650-00752-01 Rev. D...
  • Page 29: Chapter 3 Maintenance

    ZOLL AED 3 device. Self Tests The ZOLL AED 3 defibrillator performs the following self tests to verify AED integrity and its readiness for emergency use: •...
  • Page 30: Self Test Functions

    If the status indicator is blank following completion of any self test, the Self Test ZOLL AED 3 defibrillator is not ready for use and may be defective. Remove the AED from service and consult the section, “Chapter 4 Troubleshooting” on AED Failed page 29, to help determine the problem.
  • Page 31 CPR Circuitry and      Sensor (if defibrillation pads with CPR func- tionality are connected) Audio Circuitry      200 Joule Charge/    Discharge Test 9650-000752-01 Rev. D ZOLL AED 3 Administrator’s Guide...
  • Page 32: Optional Maintenance Tests

    ZOLL AED 3 models to verify that the AED is working properly and is ready for use. Note that there are two separate optional maintenance tests in this chapter: One for ZOLL AED 3 and ZOLL AED 3 BLS models, and one for the ZOLL AED 3 Automatic model.
  • Page 33: Optional Maintenance Test For Technical Professionals

    ZOLL AED 3 is working properly and is ready for use. This section has two tests: one for ZOLL AED 3 and ZOLL AED 3 BLS models, and one for ZOLL AED 3 Automatic models.
  • Page 34: Test Procedure (Zoll Aed 3 Automatic Models)

    NOTE You must use the ZOLL AED Simulator with this test. 1. Connect the ZOLL AED Simulator to the ZOLL AED 3 defibrillation pads connector. 2. Power on the simulator and the ZOLL AED 3 defibrillator. Verify that all of the following occur: •...
  • Page 35: Battery Maintenance

    Issues the CHANGE BATTERY voice prompt (if the AED is on). • Displays a blank status indicator window (no green check), indicating that • the battery is at low capacity or that the ZOLL AED 3 defibrillator has failed other self tests. Table 5. Battery Conditions Battery Icon/...
  • Page 36 ZOLL AED 3 Administrator’s Guide 9650-000752-01 Rev. D...
  • Page 37: Troubleshooting The Aed

    This chapter describes technical issues that you may encounter during routine maintenance or after a malfunction of the ZOLL AED 3 defibrillator. If this chapter does not help you resolve the problem, call the ZOLL Technical Service Department for assistance. For contact information, refer to page v.
  • Page 38 Table 6. ZOLL AED 3 Troubleshooting Problem Operator Action Technical Action Press and hold the On/Off If the AED continues to Green check does not appear in status indicator button for at least 5 seconds fail, remove it from ser- window.
  • Page 39 Table 6. ZOLL AED 3 Troubleshooting (Continued) Problem Operator Action Technical Action RELEASE SHOCK (ZOLL AED 3 and ZOLL AED 3 If the message persists, BUTTON BLS models) Release the Shock remove the AED from button. Do not press the button...
  • Page 40 ZOLL AED 3 Administrator’s Guide 9650-000752-01 Rev. D...
  • Page 41: Appendix A Specifications

    Appendix A Specifications Defibrillator Specifications This section describes product specifications for the ZOLL AED 3 defibrillator: Device Size (H x W x D) 5.0" x 9.3" x 9.7" (12.7 cm x 23.6 cm x 24.7 cm) Weight 5.5 lbs.; 2.5 kg...
  • Page 42 Defibrillator Waveform Rectilinear Biphasic™ Defibrillator Charge Hold Time ZOLL AED 3/ZOLL AED 3 BLS models: 30 seconds ZOLL AED 3 Automatic model: 3 seconds prior to automatic shock delivery Energy Selection Automatic pre-programmed selection (Adult mode: 120J, 150J, 200J; Child mode: 50J, 70J,...
  • Page 43 Wi-Fi authentication protocols. CPR Monitoring Metronome Rate: 105 ±2 CPM Compression Depth 0.75 to 4 inches ±0.25 inches 1.9 to 10.2 cm ±0.6 cm Compression Rate 50 to 150 compressions per minute 9650-000752-01 Rev. D ZOLL AED 3 Administrator’s Guide...
  • Page 44: Guidance And Manufacturer's Declaration - Electromagnetic Compatibility Guidance

    Guidance and Manufacturer’s Declaration - Electromagnetic Compatibility Guidance The ZOLL AED 3 defibrillator is intended for use in the electromagnetic environment specified below. The customer or the user of the AED should assure that it is used in such an environment.
  • Page 45 The ZOLL AED 3 defibrillator is intended for use in the electromagnetic environment specified below. The customer or the user of the AED should assure that it is used in such an environment. Immunity test IEC 60601 test Compliance Electromagnetic environment -...
  • Page 46 The ZOLL AED 3 defibrillator is suitable for use in all professional and domestic establishments. It is not intended for use near intentional transmitters of radio energy outside the ranges stated in the above table, such as high frequency surgical equipment, radar installations, or radio transmitters.
  • Page 47 Recommended separation distances between portable and mobile RF communications equipment and the ZOLL AED 3 defibrillator The ZOLL AED 3 is intended for use in an environment in which radiated RF disturbances are controlled. The customer or the user of the ZOLL AED 3 can help prevent...
  • Page 48: Rectilinear Biphasic Waveform Characteristics

    19 A 12 A 11 A Maximum Initial Current Second Phase Average 21 A 14 A 11 A 10 A Current Second Phase Duration 4 ms 4 ms 4 ms 4 ms ZOLL AED 3 Administrator’s Guide 9650-000752-01 Rev. D...
  • Page 49 Figures 1 through 6 show the Rectilinear Biphasic waveforms that are produced when the ZOLL AED 3 defibrillator is discharged into loads of 25, 50, 75, 100, 125, 150, and 175 ohms at each energy setting (200, 150, 120, 85, 70, and 50 joules).
  • Page 50 (ms). 25Ω 50Ω 75Ω 100Ω 125Ω 150Ω 175Ω Figure 1. Rectilinear Biphasic Waveforms at 200 Joules 25Ω 50Ω 75Ω 100Ω 125Ω 150Ω 175Ω Figure 2. Rectilinear Biphasic Waveforms at 150 Joules ZOLL AED 3 Administrator’s Guide 9650-000752-01 Rev. D...
  • Page 51 25Ω 50Ω 75Ω 100Ω 125Ω 150Ω 175Ω Figure 3. Rectilinear Biphasic Waveforms at 120 Joules 25Ω 50Ω 75Ω 100Ω 125Ω 150Ω 175Ω Figure 4. Rectilinear Biphasic Waveforms at 85 Joules 9650-000752-01 Rev. D ZOLL AED 3 Administrator’s Guide...
  • Page 52 25Ω 50Ω 75Ω 100Ω 125Ω 150Ω 175Ω Figure 5. Rectilinear Biphasic Waveforms at 70 Joules 25Ω 50Ω 75Ω 100Ω 125Ω 150Ω 175Ω Figure 6. Rectilinear Biphasic Waveforms at 50 Joules ZOLL AED 3 Administrator’s Guide 9650-000752-01 Rev. D...
  • Page 53: Ecg Analysis Algorithm Accuracy

    The data in the following tables summarizes the accuracy of the ECG analysis algorithm as tested against ZOLL’s ECG Rhythm Database. The algorithm sequence takes approximately six to nine seconds and proceeds as follows: Divides the ECG rhythm into three-second segments.
  • Page 54: Clinical Performance Results

    Non-shockable Specificity >99% >99% >98% AF, SB, SVT, Heart block, >95% >99% >97% idioventricular, PVCs Asystole >95% >99% >90% Intermediate Fine VF Report only >NA >NA Other VT Report only >81% >69% ZOLL AED 3 Administrator’s Guide 9650-000752-01 Rev. D...
  • Page 55 A false positive (10) is an organized or perfusing rhythm or asystole that has been incorrectly classified as a shockable rhythm. A false negative (0) is a VF or VT associated with cardiac arrest that has been incorrectly classified as non-shockable. 9650-000752-01 Rev. D ZOLL AED 3 Administrator’s Guide...
  • Page 56: Wireless Output Guidance And Manufacturer's Declaration

    Wireless Output Guidance and Manufacturer’s Declaration RF Transmission Emitted (IEC 60601-1-2) The ZOLL AED 3 unit complies with IEC 60601-1-2 for medical electrical equipment and medical electrical systems that include RF transmitters as specified below. Standard Frequency Effective Radiated Modulation...
  • Page 57: Appendix B Accessories Accessories

    Appendix B Accessories Accessories The following accessories are compatible for use with the ZOLL AED 3 defibrillator. To order any of these items, contact your local ZOLL representative. NOTE The use of accessories other than those specified in this appendix may result in increased emissions or decreased immunity of the ZOLL AED 3 defibrillator.
  • Page 58 • ILCOR Flush Wall Sign (DEA) 8000-001264 • ILCOR 3D Wall Sign (DEA) 8000-001265 Simulation/Training • ZOLL AED Simulator 8000-000925 • Training CPR Uni-padz 8900-000264 Documentation • ZOLL AED 3 Operator’s Manual 9650-000750-12 ZOLL AED 3 Administrator’s Guide 9650-000752-01 Rev. D...
  • Page 59: Appendix C Configuration Settings

    Appendix C Configuration Settings Overview This section describes the configurable settings for the ZOLL AED 3 defibrillator. The configuration settings can be manually selected through the use of the LCD touch screen, or automatically configured by reading a configuration file directly from a USB flash drive.
  • Page 60 N. A. Mountain • N. A. Central • N. A. Eastern • (default) Cuba • Columbia • Venezuela • Atlantic (includes • Puerto Rico and Virgin Islands) Paraguay • Newfoundland • ZOLL AED 3 Administrator’s Guide 9650-000752-01 Rev. D...
  • Page 61 • China • W. Australia • Japan (includes • Korea) Yakutsk • Central Australia • E. Australia • Vladivostok • Central Pacific • (includes Guam) Srednekolymsk • New Zealand • Kamchatka • 9650-000752-01 Rev. D ZOLL AED 3 Administrator’s Guide...
  • Page 62 When enabled (ON) and following the • completion of a periodic self test, the OFF (default) • ZOLL AED 3 will attempt to link to ZOLL PlusTrac via an active Wi-Fi connection. Note: The device history Wi-Fi configuration setup must be completed for this to function (see “Setting Up Device History Wi-Fi...
  • Page 63 • 70 J • 85 J (default) • Breathe During CPR When this option is enabled (ON), the • Prompt AED prompts Give Two Breaths every OFF (default) • 30 recognized compressions. 9650-000752-01 Rev. D ZOLL AED 3 Administrator’s Guide...
  • Page 64 (default) 150 seconds • 180 seconds • Wi-Fi Settings These icons contain the settings to set up your wireless connection. See “Setting Up Your Wireless Configuration” on page 57, for more information. ZOLL AED 3 Administrator’s Guide 9650-000752-01 Rev. D...
  • Page 65: Setting Up Your Wireless Configuration

    This section describes how to set up a wireless connection for your AED for uploading both clinical archives and the device history. The wireless settings are accessible when the ZOLL AED 3 defibrillator is in AED Management Mode. The Wi-Fi settings are located in the Supervisor menu for advanced users. You need a six-digit pass code to enter the Supervisor menu (see “Set Supervisor...
  • Page 66: Entering Supervisor Configuration Mode

    User ID Use the alpha-numeric keypad enter the user- name of the host server. Press OK. Note: If you have set up a ZOLL Case Review account on zollonline.com, use the ID you specified for that account. Passcode Use the alpha-numeric keypad to enter your pass- word of the host server.
  • Page 67: It Setup For Clinical Archives

    6. Press OK to save the Network settings. 7. Press the Server Settings icon to access the Server Settings window and enter information in the following fields: Server Mode: Select URL or IP. • Settings icon 9650-000752-01 Rev. D ZOLL AED 3 Administrator’s Guide...
  • Page 68 User ID - Use the alpha-numeric keypad to enter the User ID. Press OK. • NOTE If you have set up a ZOLL Case Review account on zollonline.com, use the ID you specified for that account. •...
  • Page 69: Setting Up Device History Wi-Fi Configuration

    Press OK. 2. Press the Test Wi-Fi button in the Device History Quick Settings window to test the server connectivity. NOTE If connectivity is not successful, check your wireless configuration settings. 9650-000752-01 Rev. D ZOLL AED 3 Administrator’s Guide...
  • Page 70: It Setup For Device History

    DNS table for the default value, and you will have to import a new SSL root certificate. For detailed information on how to import a root certificate, see “Importing Files from a USB Flash Drive” on page 16. ZOLL AED 3 Administrator’s Guide 9650-000752-01 Rev. D...
  • Page 71 12. Press Save to save the Wi-Fi settings. The AED displays the message SAVING CONFIGURATION, PLEASE WAIT. When the AED has saved the con- figuration, it displays the Advanced Configuration window. 9650-000752-01 Rev. D ZOLL AED 3 Administrator’s Guide...
  • Page 72 ZOLL AED 3 Administrator’s Guide 9650-000752-01 Rev. D...

This manual is also suitable for:

Aed 3 automaticAed 3 bls8503-001103-06

Table of Contents

Save PDF