AHA/ERC/ILCOR Guidelines for use of AEDs. This guide provides information about the operation and care of the ZOLL AED 3 defibrillator. The ZOLL AED 3 Administrator’s Guide is to be used by trained medical personnel in conjunction with the ZOLL AED 3 Operator’s Manual.
Appendix C—Contains definitions of all configuration settings for the ZOLL AED 3 defibrillator and provides instructions for setting up your wireless configuration. Conventions Throughout this document, text prompts and voice prompts are indicated by capital italicized letters, such as CALL FOR HELP.
• Description of the problem • If you need to send the ZOLL AED 3 defibrillator to ZOLL Medical Corporation, obtain a service order request number from the Technical Service representative. NOTE Before sending the defibrillator and battery to ZOLL, check with your local shipping company to see if there are any restrictions on shipping lithium batteries.
AED and return the AED and battery pack in its original container (or equivalent packaging) to the nearest authorized ZOLL Medical Corporation Service Center. To locate an authorized service center, contact the nearest ZOLL sales office or authorized distributor. For a list of ZOLL offices, please visit: http://www.zoll.com/contact/worldwide-locations/.
Real CPR Help for CPR monitoring and feedback. • ZOLL AED 3 BLS — Operates in Semi-Automatic Mode. Includes Real CPR Help for CPR monitoring and feedback, optional patient ECG display, CPR Dashboard, optional lay rescuer graphics, and optional audio recording.
The steps that include graphics and audio prompts are defined by local protocols for use of an AED. Child Mode The ZOLL AED 3 defibrillator has a Child mode that can be enabled in two ways: ® when the rescuer uses ZOLL CPR Uni-padz defibrillation pads and presses the ®...
If the defibrillation pads cable is not attached properly, the AED issues the PLUG IN PADS CABLE prompt. Make certain to connect the defibrillation pad cable to the ZOLL AED 3 defibrillator and attach the pads to the patient properly. WARNING! Defibrillation pads must be pre-attached to the device.
Transferring Files Using a USB Flash Drive All ZOLL AED 3 models have a USB port that allows you to export and save configuration settings, clinical archives, and device history to a USB flash drive. This feature allows you to easily copy a configuration from one AED to other AEDs.
Defibrillation Pad Connector — This connector is used to attach defibrillation pads to the AED. 9650-000752-01 Rev. D ZOLL AED 3 Administrator’s Guide...
Support Bar — The brace that opens to hold the AED in an upright position. Graphical Display Area The ZOLL AED 3 defibrillator has a 5.3 x 9.5 centimeter LCD screen that displays the following information in the modes in this section: AED Management Mode and Rescue Mode.
Rescue Mode The ZOLL AED 3 and ZOLL AED 3 Automatic models show the Lay Rescuer display, while the ZOLL AED 3 BLS model can be configured to show the Lay Rescuer display, CPR only display, or CPR and ECG display.
Page 16
CPR Only/CPR and ECG Displays (ZOLL AED 3 BLS models only) Chest compression rate CPR countdown timer Rate Depth CPR Time (in) (cpm) 0:15 Chest compression ECG rhythm depth (optional) ECG scale marker User Number of prompt GOOD COMPRESSIONS shocks...
Audio Prompts During clinical use of the ZOLL AED 3 defibrillator, you may hear the following audio prompts: Table 1. Clinical Audio Prompts Audio Prompt Definition/Action AUTOMATIC DEFIBRILLATOR The AED is the fully automatic model that has the capability to automatically shock the patient without the rescuer having to press the Shock button.
Page 18
Release the Shock Button and press again after the ready tone sounds. SHOCK WILL BE DELIVERED (ZOLL AED 3 Automatic model only) The AED will IN THREE, TWO, ONE automatically deliver shock after the countdown completes.
Page 19
Stop CPR. The AED is about to begin an ECG rhythm analysis. CPR ONLY MODE Defibrillation is not working properly; AED will only provide CPR support. Prompts that you may hear during non-clinical use of the ZOLL AED 3 defibrillator include: Table 2. Non-Clinical Audio Prompts Audio Prompt Definition...
USB flash drive. This allows you to quickly and easily update software or copy a configuration to multiple AEDs. In order to transfer data, the ZOLL AED 3 defibrillator must be in AED Management mode. To enter AED Management Mode, press and hold the On/Off button for at least 5 seconds, then press the Device Configuration icon on the main display screen.
The ZOLL AED 3 BLS model has an audio recording option that can be configured on or off. If the audio recording option is on, the ZOLL AED 3 BLS model can record and store 60 minutes of continuous audio, ECG, and clinical event data.
Press OK to return to the previous screen. NOTE If you rename a configuration file with the intention of importing it into another ZOLL AED 3 defibrillator, do not include spaces in the file name. 9650-00752-01 Rev. D ZOLL AED 3 Administrator’s Guide...
ZOLL Online. Using its Wi-Fi capability, the ZOLL AED 3 can be configured to transmit clinical event data over Wi-Fi to ZOLL Case Review. When using Case Review software to manage clinical event data, you can perform the following functions: •...
Press the Import Files icon and select one the following options to import: Configuration • • System Software Import Files • Root Certificates icon A green check displays next to the option you have selected. 9650-00752-01 Rev. D ZOLL AED 3 Administrator’s Guide...
Troubleshooting Communication Issues The following table summarizes communication messages that the ZOLL AED 3 defibrillator may issue, and their associated description and/or recommended action. If the AED is not working properly, contact ZOLL’s Technical Service Department for assistance. See “Technical Service” on page v for more information.
Page 27
THAT THE USB FLASH DRIVE IS PROPERLY INSERTED. UNABLE TO TRANSMIT The attempt to export the file to the USB REQUESTED FILE. flash drive was unsuccessful. Try the file transfer again. 9650-00752-01 Rev. D ZOLL AED 3 Administrator’s Guide...
Page 28
SYSTEM WILL RESTART. THE restarts. SCREEN WILL GO BLANK DURING THIS PROCESS. THIS MAY TAKE UP TO 5 MINUTES. DO NOT POWER OFF DURING THIS TIME. ZOLL AED 3 Administrator’s Guide 9650-00752-01 Rev. D...
ZOLL AED 3 device. Self Tests The ZOLL AED 3 defibrillator performs the following self tests to verify AED integrity and its readiness for emergency use: •...
If the status indicator is blank following completion of any self test, the Self Test ZOLL AED 3 defibrillator is not ready for use and may be defective. Remove the AED from service and consult the section, “Chapter 4 Troubleshooting” on AED Failed page 29, to help determine the problem.
Page 31
CPR Circuitry and Sensor (if defibrillation pads with CPR func- tionality are connected) Audio Circuitry 200 Joule Charge/ Discharge Test 9650-000752-01 Rev. D ZOLL AED 3 Administrator’s Guide...
ZOLL AED 3 models to verify that the AED is working properly and is ready for use. Note that there are two separate optional maintenance tests in this chapter: One for ZOLL AED 3 and ZOLL AED 3 BLS models, and one for the ZOLL AED 3 Automatic model.
ZOLL AED 3 is working properly and is ready for use. This section has two tests: one for ZOLL AED 3 and ZOLL AED 3 BLS models, and one for ZOLL AED 3 Automatic models.
NOTE You must use the ZOLL AED Simulator with this test. 1. Connect the ZOLL AED Simulator to the ZOLL AED 3 defibrillation pads connector. 2. Power on the simulator and the ZOLL AED 3 defibrillator. Verify that all of the following occur: •...
Issues the CHANGE BATTERY voice prompt (if the AED is on). • Displays a blank status indicator window (no green check), indicating that • the battery is at low capacity or that the ZOLL AED 3 defibrillator has failed other self tests. Table 5. Battery Conditions Battery Icon/...
This chapter describes technical issues that you may encounter during routine maintenance or after a malfunction of the ZOLL AED 3 defibrillator. If this chapter does not help you resolve the problem, call the ZOLL Technical Service Department for assistance. For contact information, refer to page v.
Page 38
Table 6. ZOLL AED 3 Troubleshooting Problem Operator Action Technical Action Press and hold the On/Off If the AED continues to Green check does not appear in status indicator button for at least 5 seconds fail, remove it from ser- window.
Page 39
Table 6. ZOLL AED 3 Troubleshooting (Continued) Problem Operator Action Technical Action RELEASE SHOCK (ZOLL AED 3 and ZOLL AED 3 If the message persists, BUTTON BLS models) Release the Shock remove the AED from button. Do not press the button...
Appendix A Specifications Defibrillator Specifications This section describes product specifications for the ZOLL AED 3 defibrillator: Device Size (H x W x D) 5.0" x 9.3" x 9.7" (12.7 cm x 23.6 cm x 24.7 cm) Weight 5.5 lbs.; 2.5 kg...
Guidance and Manufacturer’s Declaration - Electromagnetic Compatibility Guidance The ZOLL AED 3 defibrillator is intended for use in the electromagnetic environment specified below. The customer or the user of the AED should assure that it is used in such an environment.
Page 45
The ZOLL AED 3 defibrillator is intended for use in the electromagnetic environment specified below. The customer or the user of the AED should assure that it is used in such an environment. Immunity test IEC 60601 test Compliance Electromagnetic environment -...
Page 46
The ZOLL AED 3 defibrillator is suitable for use in all professional and domestic establishments. It is not intended for use near intentional transmitters of radio energy outside the ranges stated in the above table, such as high frequency surgical equipment, radar installations, or radio transmitters.
Page 47
Recommended separation distances between portable and mobile RF communications equipment and the ZOLL AED 3 defibrillator The ZOLL AED 3 is intended for use in an environment in which radiated RF disturbances are controlled. The customer or the user of the ZOLL AED 3 can help prevent...
19 A 12 A 11 A Maximum Initial Current Second Phase Average 21 A 14 A 11 A 10 A Current Second Phase Duration 4 ms 4 ms 4 ms 4 ms ZOLL AED 3 Administrator’s Guide 9650-000752-01 Rev. D...
Page 49
Figures 1 through 6 show the Rectilinear Biphasic waveforms that are produced when the ZOLL AED 3 defibrillator is discharged into loads of 25, 50, 75, 100, 125, 150, and 175 ohms at each energy setting (200, 150, 120, 85, 70, and 50 joules).
The data in the following tables summarizes the accuracy of the ECG analysis algorithm as tested against ZOLL’s ECG Rhythm Database. The algorithm sequence takes approximately six to nine seconds and proceeds as follows: Divides the ECG rhythm into three-second segments.
Non-shockable Specificity >99% >99% >98% AF, SB, SVT, Heart block, >95% >99% >97% idioventricular, PVCs Asystole >95% >99% >90% Intermediate Fine VF Report only >NA >NA Other VT Report only >81% >69% ZOLL AED 3 Administrator’s Guide 9650-000752-01 Rev. D...
Page 55
A false positive (10) is an organized or perfusing rhythm or asystole that has been incorrectly classified as a shockable rhythm. A false negative (0) is a VF or VT associated with cardiac arrest that has been incorrectly classified as non-shockable. 9650-000752-01 Rev. D ZOLL AED 3 Administrator’s Guide...
Wireless Output Guidance and Manufacturer’s Declaration RF Transmission Emitted (IEC 60601-1-2) The ZOLL AED 3 unit complies with IEC 60601-1-2 for medical electrical equipment and medical electrical systems that include RF transmitters as specified below. Standard Frequency Effective Radiated Modulation...
Appendix B Accessories Accessories The following accessories are compatible for use with the ZOLL AED 3 defibrillator. To order any of these items, contact your local ZOLL representative. NOTE The use of accessories other than those specified in this appendix may result in increased emissions or decreased immunity of the ZOLL AED 3 defibrillator.
Appendix C Configuration Settings Overview This section describes the configurable settings for the ZOLL AED 3 defibrillator. The configuration settings can be manually selected through the use of the LCD touch screen, or automatically configured by reading a configuration file directly from a USB flash drive.
Page 60
N. A. Mountain • N. A. Central • N. A. Eastern • (default) Cuba • Columbia • Venezuela • Atlantic (includes • Puerto Rico and Virgin Islands) Paraguay • Newfoundland • ZOLL AED 3 Administrator’s Guide 9650-000752-01 Rev. D...
Page 61
• China • W. Australia • Japan (includes • Korea) Yakutsk • Central Australia • E. Australia • Vladivostok • Central Pacific • (includes Guam) Srednekolymsk • New Zealand • Kamchatka • 9650-000752-01 Rev. D ZOLL AED 3 Administrator’s Guide...
Page 62
When enabled (ON) and following the • completion of a periodic self test, the OFF (default) • ZOLL AED 3 will attempt to link to ZOLL PlusTrac via an active Wi-Fi connection. Note: The device history Wi-Fi configuration setup must be completed for this to function (see “Setting Up Device History Wi-Fi...
Page 63
• 70 J • 85 J (default) • Breathe During CPR When this option is enabled (ON), the • Prompt AED prompts Give Two Breaths every OFF (default) • 30 recognized compressions. 9650-000752-01 Rev. D ZOLL AED 3 Administrator’s Guide...
Page 64
(default) 150 seconds • 180 seconds • Wi-Fi Settings These icons contain the settings to set up your wireless connection. See “Setting Up Your Wireless Configuration” on page 57, for more information. ZOLL AED 3 Administrator’s Guide 9650-000752-01 Rev. D...
This section describes how to set up a wireless connection for your AED for uploading both clinical archives and the device history. The wireless settings are accessible when the ZOLL AED 3 defibrillator is in AED Management Mode. The Wi-Fi settings are located in the Supervisor menu for advanced users. You need a six-digit pass code to enter the Supervisor menu (see “Set Supervisor...
User ID Use the alpha-numeric keypad enter the user- name of the host server. Press OK. Note: If you have set up a ZOLL Case Review account on zollonline.com, use the ID you specified for that account. Passcode Use the alpha-numeric keypad to enter your pass- word of the host server.
6. Press OK to save the Network settings. 7. Press the Server Settings icon to access the Server Settings window and enter information in the following fields: Server Mode: Select URL or IP. • Settings icon 9650-000752-01 Rev. D ZOLL AED 3 Administrator’s Guide...
Page 68
User ID - Use the alpha-numeric keypad to enter the User ID. Press OK. • NOTE If you have set up a ZOLL Case Review account on zollonline.com, use the ID you specified for that account. •...
Press OK. 2. Press the Test Wi-Fi button in the Device History Quick Settings window to test the server connectivity. NOTE If connectivity is not successful, check your wireless configuration settings. 9650-000752-01 Rev. D ZOLL AED 3 Administrator’s Guide...
DNS table for the default value, and you will have to import a new SSL root certificate. For detailed information on how to import a root certificate, see “Importing Files from a USB Flash Drive” on page 16. ZOLL AED 3 Administrator’s Guide 9650-000752-01 Rev. D...
Page 71
12. Press Save to save the Wi-Fi settings. The AED displays the message SAVING CONFIGURATION, PLEASE WAIT. When the AED has saved the con- figuration, it displays the Advanced Configuration window. 9650-000752-01 Rev. D ZOLL AED 3 Administrator’s Guide...