Setting Upper and Lower EtCO
Initially, the EtCO2 Alarm menu specifies that the EtCO
(
), and displays the default Upper and Lower EtCO
OFF
be
or
ON
pediatric, and neonate patients, and gives the range in which you can set these limits:
Patient Type
Adult
Pediatric
Neonate
Caution
In high-altitude environments, EtCO
described by Dalton's law of partial pressures. When using the X Series unit in high-altitude
environments, it is advisable to adjust EtCO
Setting Upper FiCO
Initially, the FiCO2 Alarm menu specifies that the FiCO
(
), and displays the default upper limit. The upper limit can be
OFF
The following table lists the default FiCO
patients, and gives the range in which you can set these limits:
Patient Type
Adult
Pediatric
Neonate
9650-001355-01 Rev. M
(default is
). The following table lists the default EtCO
OFF
OFF
EtCO
Limit Default
2
Lower: 8 mmHg
Upper: 60 mmHg
Lower: 8 mmHg
Upper: 60 mmHg
Lower: 8 mmHg
Upper: 60 mmHg
Limits
2
Upper FiCO
2
Default
8 mmHg
8 mmHg
8 mmHg
X Series Operator's Guide
Setting CO2 and Respiration Rate Alarms
Limits
2
alarms are enabled (
2
Limits. The upper and lower limits can
2
EtCO
Limit Range
2
Lower: 0-145 mmHg
Upper: 5-150 mmHg
Lower: 0-145 mmHg
Upper: 5-150 mmHg
Lower: 0-145 mmHg
Upper: 5-150 mmHg
values may be lower than values observed at sea level, as
2
alarm settings accordingly.
2
alarms are enabled (
2
upper limits for adult, pediatric, and neonate
2
Limit
Upper FiCO
Limit
2
Range
2-98 mmHg
2-98 mmHg
2-98 mmHg
) or disabled
ON
limits for adult,
2
) or disabled
ON
or
(default is
ON
OFF
OFF
9-9
).