Warnings -- SpO
As with all medical equipment, carefully route patient cabling to reduce the possibility of
•
patient entanglement or strangulation.
Do not place the pulse co-oximeter or accessories in any position that might cause it to fall
•
on the patient.
Do not start or operate the pulse co-oximeter unless the setup was verified to be correct.
•
Do not use the pulse co-oximeter during magnetic resonance imaging (MRI) or in an MRI
•
environment.
Do not use the pulse co-oximeter if it appears or is suspected to be damaged.
•
Explosion hazard: Do not use the pulse co-oximeter in the presence of flammable
•
anesthetics or other flammable substance in combination with air, oxygen-enriched
environments, or nitrous oxide.
To ensure safety, avoid stacking multiple devices or placing anything on the instrument
•
during operation.
To protect against injury, follow the directions below:
•
Avoid placing the device on surfaces with visible liquid spills.
•
Do not soak or immerse the device in liquids.
•
Do not attempt to sterilize the device.
•
Use cleaning solutions only as instructed in this operator's manual.
•
Do not attempt to clean the device while monitoring patient.
•
To protect from electric shock, always remove the sensor and completely disconnect the
•
pulse co-oximeter before bathing the patient.
If any measurement seems questionable, first check the patient's vital signs by alternate
•
means and then check the pulse co-oximeter for proper functioning.
Inaccurate respiration rate measurements may be caused by:
•
Improper sensor application
•
Low arterial perfusion
•
Motion artifact
•
Low arterial oxygen saturation
•
Excessive ambient or environmental noise
•
Inaccurate SpCO and SpMet readings can be caused by:
•
Improper sensor application
•
Intravascular dyes such as indocyanine green or methylene blue
•
Abnormal hemoglobin levels
•
Low arterial perfusion
•
Low arterial oxygen saturation levels including altitude induced hypoxemia
•
Elevated total bilirubin levels
•
Motion artifact
•
9650-001355-01 Rev. M
General
2
X Series Operator's Guide
Warnings -- SpO2 General