CONFIGURATION GUIDE
9650-0204-01 Rev A
October 2006

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the M Series CCT and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for ZOLL M Series CCT

  • Page 1 CONFIGURATION GUIDE 9650-0204-01 Rev A October 2006...
  • Page 2 An issue or revision date for this manual is shown on the front cover. If more than three years have elapsed since this date, contact ZOLL Medical Corporation to determine if additional product information updates are available. ZOLL, M Series, and stat padz are registered trademarks, and PD, and XL are trademarks of ZOLL Medical Corporation.
  • Page 3: Table Of Contents

    Table of Contents INTRODUCTION..........................1 CONFIGURING THE UNIT ......................1 ELECT ANGUAGE ....................3 ONFIGURATION ..........................3 RINT ONFIG ........................3 ESTORE EFAULTS ........................4 ........................5 HANGE ONFIG CONFIGURATION SETTINGS ......................7 ..........................7 OTCH ILTER .........................7 LLOW RASE ......................7 ISPLAY LAPSED ......................7 OICE ARKERS NABLED ....................8 EPORT ESTART ELAYS...
  • Page 4 ’ ERIES PERATOR UIDE ......................18 ROUP EFAULT 12 L ....................18 RINT NTERPRETATION ......................18 EFIB EFAULT TO ........................19 HONE CO2 O ........................19 PTIONS .........................19 NABLE EADS ......................19 LOOD RESSURE NITS NIBP A .....................19 NTERVAL EFAULT NIBP M BP A ..................19 RIGGER EAS ON LARM NIBP M...
  • Page 5: Introduction

    Introduction The M Series CCT is designed with several user selectable features that allow operators to configure the device according to their protocols and local requirements. This manual describes the configuration options and how to change them. This manual also lists the configurable features, options and default settings.
  • Page 6: Select Language

    Language option. Press the Prev Value and Next Value softkeys to move the highlight from one language to the other. Some M Series CCT units include software that allows the device to operate in either of two languages. If the languages displayed do not meet your needs, contact ZOLL Technical Service for a language upgrade.
  • Page 7: Main Configuration Menu

    Configuring the Unit Main Configuration Menu The Main Configuration menu is displayed whenever the Enter softkey is depressed from the Language Select menu or the Main Menu softkey is depressed while displaying other system configuration screens. This menu provides the following five (5) softkey options: Print Config, Restore Defaults, Read From Card, Change Config, and Exit Config.
  • Page 8: Read From Card

    CCT C ERIES ONFIGURATION UIDE Read From Card Pressing the Read From Card softkey prompts the user to choose either a language or configuration card. The following screen will be displayed. To move the highlighted area from LANGUAGE to CONFIGURATION and vice versa, press the Prev Item and Next Item softkeys.
  • Page 9: Change Config

    Configuring the Unit Change Config Pressing the Change Config softkey causes the unit to display the first configuration option menu. There are three user configurable features contained within each option menu. (Refer to the “Configuration Option Tables” on page 29 for all available options and default settings).
  • Page 10 CCT C ERIES ONFIGURATION UIDE (This page has intentionally been left blank.)
  • Page 11: Configuration Settings

    Sets the ECG Notch Filter frequency for proper AC mains interference rejection. Allow Card Erase Allows the user to erase PCMCIA data cards on the M Series CCT unit when set to “Yes”. Display Elapsed Time Allows the user to display the elapsed time since the unit was turned on, when enabled. The elapsed time will continue to be counted for up to 10 seconds after power down.
  • Page 12: Set Report Restart Delays

    CCT C ERIES ONFIGURATION UIDE Set Report Restart Delays Allows review of, or changes to the length of time the device must be turned off before it erases all data stored in either summary report memory or trend memory. With Set Report Restart Delay highlighted and Set/Review selected, press the Change Value softkey.
  • Page 13 Configuration Settings With Code Markers highlighted and Config/Review selected, press the Change Value softkey. The following screen displays: Pressing the Change Value softkey will change the highlighted setting from Change/Review to Set to Defaults and vice versa. Pressing the Prev Item and Next Item softkeys will scroll the highlight to the next or previously highlighted code markers feature.
  • Page 14: Device Identifier

    CCT C ERIES ONFIGURATION UIDE Device Identifier Sets an eleven-digit device identifier code. This code is divided into two sub codes (Site ID, eight alphanumeric characters, and Device ID, three alphanumeric characters). The device ™ ® identifier is printed on the 12-lead strip chart, summary report, fax and Catalyst MUSE printouts.
  • Page 15: Manual Mode Passcode (Advanced Aed Models)

    Configuration Settings Manual Mode Passcode (Advanced AED Models) Allows the three-digit Manual Mode Passcode to be set or changed. Each digit entered can only be set between 0 (zero) and 3 (three) inclusive. Entering “000” as the pass code disables password protection of the Manual Mode. When “000” is selected, no password is required to enter the Manual Defibrillation Mode.
  • Page 16: Energy Level : Shock 1, 2, 3

    CCT C ERIES ONFIGURATION UIDE Energy Level: Shock 1, 2, 3 Selects the energy levels used for the 1 , and 3 consecutive shocks that the defibrillator delivers when: In semi-automatic or advisory mode and using pads • Ιn manual mode, using external paddles or pads, with Basic Auto Energy Escalation set •...
  • Page 17: Store To Card In Auto Monitor Mode (Aed Models)

    CCT unit is on M Seri • If disabled, the M Series CCT unit will not record any data to the PCMCIA card unless MFE pads are connected to the patient. Auto Charge in Advisory Mode When enabled and a shockable rhythm is detected after pressing the ANALYZE button, the unit will automatically charge the defibrillator to the pre-configured or user selected energy setting.
  • Page 18: Pace Rate Setting On Power-Up

    ONFIGURATION UIDE Pace Rate Setting on Power-Up Sets the default pace rate of the M Series CCT unit on power-up, between 30 ppm and 102 ppm (increments of 4 ppm). Display Implanted Pace Spikes The M Series CCT is capable of detecting and indicating on the display pacemaker signals from a patient with an implantable pacemaker.
  • Page 19: Fax/Communication Phone Numbers

    Fax/Communication Phone Numbers Allows the user to pre-configure up to 24 locations and phone numbers and delete obsolete phone numbers. The M Series CCT uses the phone numbers to transmit data to a fax machine or Catalyst MUSE system following 12-lead analysis.
  • Page 20 CCT C ERIES ONFIGURATION UIDE 4. Press Details softkey to select field. The Location and Phone Number menu displays: To update location of medical facility: 1. Press Prev Letter or Next Letter softkey to select letter. 2. Press Inc Letter or Dec Letter softkey to change value of letter. Repeat steps 1 and 2 until location has been entered.
  • Page 21: Lead Analysis

    Configuration Settings Deleting Fax/Communication Information The following procedure deletes obsolete phone numbers from your M Series CCT unit. To delete phone numbers: 1. From the main System Configuration menu, press Next Item softkey until “FAX/ COMMUNICATION PHONE NUMBERS” highlights. 2. Press Change Value softkey.
  • Page 22: Set Lead Groups

    CCT C ERIES ONFIGURATION UIDE Set Lead Groups This setting allows the user to configure two custom lead groups as described in the 12-Lead ECG Monitoring insert (part number 9650-0215-01). Pressing the Change Value softkey will cause the following screen to appear: Pressing the Prev Lead softkey will select the previous lead.
  • Page 23: Unit:

    When set to “Yes” and ECG leads do not have proper contact with the patient, or the cable is not properly attached to the M Series CCT unit, the ECG LEADS OFF message displays and a dashed line displays in place of the ECG waveform. Yes is the default setting.
  • Page 24: E Scalation

    ONFIGURATION UIDE On M Series CCT units, you can set the NIBP cuff inflation pressure for three patient type modes: adult, pediatric and neonatal. With NIBP First Inflate Pressure Default highlighted and Set/Review selected, press the Change Value softkey. The following screen displays: Pressing the Change Value softkey will scroll through the available inflation pressures for the highlighted parameter.
  • Page 25: L Ocation

    M Series CCT and stored in memory once each minute. When displaying this data in tabular format on the M Series CCT screen, however, the user can choose to view all or only a subset of the stored trend data. The factory default zoom level causes all one-minute data samples to be displayed when viewing the trend table.
  • Page 26 5, 10, 15, 30, or 60 minutes. All data samples are printed on the strip chart regardless of this setting. Trending Enabled When set to “Yes”, the M Series CCT unit will print and display trend reports as long as the appropriate hardware is installed. Defaults to “No” for units not equipped with the appropriate hardware.
  • Page 27: Power Up

    Sets the interval between Low Battery warning beeps to either 1 minute or 5 minutes. The Low Battery warning message is displayed once per minute, regardless of this setting. Clock Synchronization Allows the user to set clock synchronization parameters on the M Series CCT unit.
  • Page 28: M Essage E Nabled

    Erase Card Prompt at Pwr Off Sets the M Series CCT unit to allow the user to erase the data card at power-off if the card contains only self test, and optionally, synchronized cardioversion data. To enable this option,...
  • Page 29: T Ype (Nibp)

    Remove Card Prompt at Pwr Off Sets the M Series CCT unit to prompt the user to remove the data card at power-off if the card contains any information other than self test, and optionally, synchronized cardioversion data.
  • Page 30: Nits

    Sets the temperature measurement unit to °F or °C. Display Colors Allows the user to set individual display colors for parameters displayed on the M Series CCT monitor. Highlight the Display Colors option and press the Change Value softkey to display the following screen: Press the Change Value softkey to scroll through the available colors for each parameter.
  • Page 31: Power Up

    Configuration Settings IBP1 Label at Power Up Sets the label assigned to the first IBP channel when the unit is powered up. IBP2 Label at Power Up Sets the label assigned to the second IBP channel when the unit is powered up. IBP1 Range at Power Up Sets the default display scale (in mmHg) for the waveform amplitude of the first IBP channel at power-up.
  • Page 32 CCT C ERIES ONFIGURATION UIDE (This page has intentionally been left blank.)
  • Page 33: Configuration Option Tables

    The following tables list the M Series CCT options and their possible values. An asterisk (“*”) refers to features that are currently not implemented on the M Series CCT. These features will only operate in accordance with their pre-configured default settings (listed below).
  • Page 34 CCT C ERIES ONFIGURATION UIDE Feature Options Default Clock Synchronization Set/Review Set/Review Time Zone Azores, Middle Atlantic, S. America Eastern, EST Eastern AST Atlantic, EST Eastern, CST Central, MST Mountain, PST Pacific, AKST Alaska, HST Hawaiian, Samoa, Dateline, GMT Greenwich, CET Central Europe, EET Eastern Europe, MSK Moscow, Arabian, Afghanistan, West Asia, Central Asia, Southeast Asia, AWST Australian...
  • Page 35 Configuration Option Tables Advisory and Semi-automatic Modes Feature Options Default Energy Level: Shock 1 Damped Sine Wave (DSW): 1-10, 15, 20, 30, Manual DSW: 200 J 50, 75, 100, 150, 200, 300, 360 J Manual Biphasic: 120 J Biphasic: 1-10, 15, 20, 30, 50, 75, 100, 120, AED DSW: 200 J 150, 200 J AED Biphasic: 120 J...
  • Page 36 CCT C ERIES ONFIGURATION UIDE Manual Mode Feature Options Default Auto Generate Strips Yes/No Retain SYNC after Defib Yes/No Print 3 Leads When Leads Sel. Yes/No Auto Transmit After 12 Lead Yes/No Fax/Communication Phone Numbers Set/Review Set/Review 12 Lead Analysis Current 10 sec Current 10 sec 12 Lead Printout...
  • Page 37: Option

    Configuration Option Tables EtCO Option Feature Options Default EtCO Options Set/Review Set/Review Default EtCO Units mmHg, kPa, % mmHg Displayed Zoom Level (in mmHg): 0-12.5, 0-25, 0-50, 0-75, 0-100 0-50 mmHg (in kPa): 0-1.7, 0-3.3, 0-6.6, 0-10.0, 0-13.3 0-6.6 kPa Default EtCO Comp Setting None, O2, N2O, O2 &...
  • Page 38: Ption

    CCT C ERIES ONFIGURATION UIDE IBP Option Feature Options Default Blood Pressure Units mmHg, kPa mmHg IBP1 Label at Power Up P1, ART, PA, CVP, ICP IBP2 Label at Power Up P2, ART, PA, CVP, ICP IBP1 Range at Power Up (in mmHg): 0-300, 0-180, 0-120, 0-60, 0-30 0-300 mmHg (in kPa): 0-40, 0-24, 0-16, 0-8, 0-4...