Setting Alarms

Before taking a measurement, ensure that all the NIBP alarms settings are appropriate for the
situation. See the M Series Operator's Guide for instructions on setting alarms. When an NIBP or
heart rate alarm is triggered, the unit automatically initiates a single blood pressure measurement. See
the M Series Configuration Guide for instructions on how to enable/disable this automatic
measurement.
The M Series NIBP option includes blood pressure alarms for the following measurements:
Table 4 Alarm Limit Settings and Defaults
Patient Type
Adult
Pediatric*
Neonate*
* Neonatal and pediatric NIBP modes available on M Series CCT units only.
When the M Series unit is turned on, all alarm functions are disabled and the alarm limits are set to
their default values. Any changes to these settings remain in effect until either the settings are changed
or 10 seconds after the M Series unit is turned off. On M Series CCT units, you can configure the
default alarm limits at power-up according to patient type; see the M Series Configuration Guide for
more information.
You can change all the alarm limits using standard M Series procedures. See the M Series Operator's
Setting Alarm
Limits
Guide for instructions on setting alarm limits.
You can set the alarm limits using the Auto function, or you can choose your own limits (see Table
4 on page 11). The Auto function sets the high limit to 30 mmHg (4.0 kPa) above and the low limit to
20 mmHg (2.7 kPa) below the last measurement.
Alarm Parameter
Default Setting
Systolic
High
160 mmHg (21.3 kPa)
Low
90 mmHg (12.0 kPa)
Diastolic
High
110 mmHg (14.7 kPa)
Low
50 mmHg (6.7 kPa)
Mean
High
130 mmHg (17.3 kPa)
Low
60 mmHg (8.0 kPa)
Systolic
High
145 mmHg (19.3 kPa)
Low
75 mmHg (10.0 kPa)
Diastolic
High
100 mmHg (13.3 kPa)
Low
35 mmHg (4.7 kPa)
Mean
High
110 mmHg (14.7 kPa)
Low
50 mmHg (6.7 kPa)
Systolic
High
100 mmHg (13.3 kPa)
Low
50 mmHg (6.7 kPa)
Diastolic
High
70 mmHg (9.3 kPa)
Low
30 mmHg (4.0 kPa)
Mean
High
80 mmHg (10.7 kPa)
Low
35 mmHg (4.7 kPa)
ZOLL M Series Operator's Guide - Option Insert
Preparing to Take Measurements
Range (increments of 5)
80-260 mmHg (10.7-34.7 kPa)
40-140 mmHg (5.3-18.7 kPa)
50-200 mmHg (6.7-26.7 kPa)
25-100 mmHg (3.3-13.3 kPa)
60-220 mmHg (8.0-29.3 kPa)
30-120 mmHg (4.0-16.0 kPa)
80-160 mmHg (10.7-21.3 kPa)
35-140 mmHg (4.7-18.7 kPa)
50-130 mmHg (6.7-17.3 kPa)
20-100 mmHg (2.7-13.3 kPa)
60-140 mmHg (8.0-18.7 kPa)
20-120 mmHg (2.7-16.0 kPa)
60-130 mmHg (8.0-17.3 kPa)
25-120 mmHg (3.3-16.0 kPa)
30-105 mmHg (4.0-14.0 kPa)
20-100 mmHg (2.7-13.3 kPa)
35-110 mmHg (4.7-14.7 kPa)
20-105 mmHg (2.7-14.0 kPa)