Heart Rate Meter

The Heart Rate meter displays the QRS complex rate that it derives from the ECG monitoring
function. The Heart Rate meter always computes the heart rate from the top waveform. The
Heart Rate meter is labeled HR (as in the following example) if the source is ECG, and PR if
SPO2 is measured and ECG cable is disconnected.

Configuring Heart Rate (HR) Meter Alarms

The ZOLL M2 unit allows you to enable and disable the Heart Rate (HR) alarm, to set alarm
limits, and to select a QRS detection tone volume. The default HR alarm settings (enable/
disable, alarm limits) are supervisor configurable.
Heart Rate (HR) Alarm Limits
Initially, the HR Alarm Settings menu specifies that alarms are enabled (
and displays the default Upper and Lower heart rate alarm limits. The following table lists the
default HR alarm limits for adult and pediatric patients, and gives the range in which you can
set these limits:
Patient Type
Adult/Pediatric
9650-000860-01 Rev. C
HR Default Limit Alarm
Lower: 50 BPM
Upper: 120 BPM
ZOLL M2 Operator's Guide
Heart Rate Meter
or disabled (
)
ON
HR Alarm Limits Range
Lower: 20 to 299 BPM
Upper: 21 to 300 BPM
Heart Rate
Meter
),
OFF
6-11