Setting Alar ms
Before taking measurements, check that all IBP alarms settings are appropriate for the situation. You
can set the alarms for Pressure channel 1, Pressure channel 2 and NIBP separately. See M Series
Operator's Guide for instructions on setting alarms.
C
!
Before the M Series with IBP option is initially put into service, the IBP alarm limit
AUTION
defaults should be configured per institutional guidelines. See the M Series CCT
Configuration Guide for further details.
The M Series IBP option includes blood pressure alarms for the following measurements:
Table 1: Alarm Limit Settings and Defaults - Arterial Pressure (ART)
Alarm Limit:
High Systolic
Low Systolic
High Diastolic
Low Diastolic
High Mean
Low Mean
Table 2: Alarm Limit Settings and Defaults - Pulmonary Arterial Pressure (PA)
Alarm Limit:
High Systolic
Low Systolic
High Diastolic
Low Diastolic
High Mean
Low Mean
Table 3: Alarm Limit Settings and Defaults - Central Venous Pressure (CVP)
Alarm Limit:
High
Low
Table 4: Alarm Limit Settings and Defaults - Intracranial Pressure (ICP)
Alarm Limit:
High
Default Setting:
160 mmHg; 21.3 kPa
90 mmHg; 12.0 kPa
110 mmHg; 14.6 kPa
50 mmHg; 6.6 kPa
130 mmHg; 17.3 kPa
60 mmHg; 8.0 kPa
Default Setting:
300 mmHg; 39.9 kPa
-50 mmHg; -6.7 kPa
300 mmHg; 39.9 kPa
-50 mmHg; -6.7 kPa
300 mmHg; 39.9 kPa
-50 mmHg; -6.7 kPa
Default Setting:
300 mmHg; 39.9 kPa
-50 mmHg; -6.7 kPa
Default Setting:
300 mmHg; 39.9 kPa
ZOLL M Series Invasive Blood Pressure
Range (increments of 5):
-45 to 300 mmHg; -6.0 to 39.9 kPa
-50 to 295 mmHg; -6.7 to 39.2 kPa
-45 to 300 mmHg; -6.0 to 39.9 kPa
-50 to 295 mmHg;-6.7 to 39.2 kPa
-45 to 300 mmHg; -6.0 to 39.9 kPa
-50 to 295 mmHg; -6.7 to 39.2 kPa
Range (increments of 5):
-45 to 300 mmHg; -6.0 to 39.9 kPa
-50 to 295 mmHg; -6.7 to 39.2 kPa
-45 to 300 mmHg; -6.0 to 39.9 kPa
-50 to 295 mmHg;-6.7 to 39.2 kPa
-45 to 300 mmHg; -6.0 to 39.9 kPa
-50 to 295 mmHg; -6.7 to 39.2 kPa
Range (increments of 5):
-45 to 300 mmHg; -6.0 to 39.9 kPa
-50 to 295 mmHg; -6.7 to 39.2 kPa
Range (increments of 5):
-45 to 300 mmHg; -6.0 to 39.9 kPa
11