Pace Rate Setting on Power-Up

Sets the default pace rate of the M Series CCT unit on power-up, between 30 ppm and 102 ppm
(increments of 4 ppm).

Display Implanted Pace Spikes

The M Series CCT is capable of detecting and indicating on the display pacemaker signals from
a patient with an implantable pacemaker. Selecting "Yes" causes a pacemaker marker to be
displayed on the ECG trace whenever implanted pacer stimuli are detected. The pacemaker
spike detector is always active in ECG leads, independent of this configuration setting. This
setting controls spike display only.

Gain On Power-Up

Sets the default SIZE of the displayed ECG signal (AGC, 0.5 cm/mV, 1 cm/mV, 1.5 cm/mV,
and 2 cm/mV) when the device is powered-up.

Retain SYNC after Defib

Causes the unit to remain in SYNC mode after a synchronized cardioversion shock. The unit
will remain in SYNC mode until the SYNC button is pressed again or the unit is switched out of
Defib mode.

Auto Generate Strips

Sets the strip chart recorder to automatically begin printing after a defibrillator discharge or
whenever alarms are triggered while in manual mode. (Unit must be equipped with the optional
strip chart recorder feature.)

Print 3 Leads When Leads Sel

When enabled, the strip chart recorder will automatically print three (3) simultaneous leads of
the patient ECG when Leads are selected and a five (5) lead or 12-lead ECG cable is in use.

Report Memory Card Errors

Allows the user to disable memory card error messages and voice prompts.
If this option is set to "No", all error messages (i.e., "INSERT CARD", "REPLACE CARD",
"CARD FULL", and "CARD LOW < n MINUTES") related to the use of the memory card will
be suppressed. However, if the unit is in manual mode, or the unit is non-AED, the option is
ignored when the "Record to Card in Advisory Mode" option is set to "None" (no data
recording).

Auto Analyze on Power Up (AED Semi-Automatic Mode)

When set to "Yes", the unit will automatically begin an ECG rhythm analysis as soon as the
device is turned on and pads are properly connected. An "ANALYZING ECG" message will be
displayed for 5 seconds and a "STAND CLEAR" message is displayed and announced
indicating the start of the analysis. If the Auto Analyze 3 Times option is enabled, the analysis
will be the first of a three analysis sequence.
Note:

Auto Transmit After 12 Lead

When enabled, the unit will automatically enter the transmission setup screen following
completion of a 12-lead analysis.
To fax using 2x6 format, this must be set to "Yes."
G
UIDE
The unit must be turned off for 10 seconds or more for the analysis to auto -start when
the device is turned on.
14