Replacing The I.i; Checking The Temperature Indicator; Removal - Siemens SIREMOBIL Iso-C Troubleshooting Manual

Table of Contents

Advertisement

Replacing boards and components

Replacing the I.I.

Checking the temperature indicator

Prior to installing the I.I., check the temperature indicator.
If the inner square field of the indicator is white, the I.I. did not exceed the temperature
range. Remove the temperature indicator.
If the indicator is discolored (inner field black), proceed according to IQ document
AX51-030.812.01...

Removal

WARNING
WARNING
CAUTION
VIDEOMED DC
Move the I.I. into the service position. See
Remove the Videomed DC and place it on a clean surface.
Cover it to prevent contamination. See also
© Siemens, 2008
For internal use only
Electrical voltage!
See
(Safety information / p.
Switch the SIREMOBIL Iso-C off and wait at least > 3
minutes until the high voltage in the system and in the
I.I. mini-voltage supply dissipates.
Risk of injury on mechanical parts!
If these guidelines are not followed, minor to moderate injury,
especially to the hands, can occur.
Follow the notes concerning the C-arm carbon fiber
structure.
Risk of injury!
Noncompliance can lead to minor to medium injury and/or
property damage.
The C-arm covers contain counterbalance weights and weigh
approx. 6 kg.
Hold the covers firmly when installing or removing them.
Before the C-arm cover is loosened, the C-arm must be
moved to its middle orbital position. If the C-arm cov-
ers are loosened, the C-arm may not be moved to its
orbital end positions (I.I. or SIREPHOS at the orbital
end position).
SPR2-230.840.01.01.02
08.08
11).
(I.I. service position / p.
(Replacing the VIDEOMED DC / p.
Page 95 of 148