Test 5: Phantom Imaging Quality
Objective
To ensure that adequate phantom imaging quality is achieved
Required Equipment
a)
ACR (RMI 156) phantom
b)
Compression plate 24 x 30 cm
Procedure
1.
Create a new patient record in the local database.
Fill in: Last Name:
2.
Select procedure QC-raw.
3.
Select exposure technique AEC
28 kV
target/filter combination Mo/Mo
grid in
auto decompression off
OPCOMP off
AEC segmentation on
low dose mode off
Change all projection views to L-CC
MAMMOMAT Inspiration
SPB7-330.640.01.01.02
Test_Five
First Name: Phantom Imaging Quality
Patient ID:
day+time when the test is performed
(example: 200707071030)
DOB:
05 05 1955
Gender:
Other
Test 5: Phantom Imaging Quality