Otto Bock Skippi Plus Service Manual page 64

Hide thumbs Also See for Skippi Plus:
Table of Contents

Advertisement

enAble50 wheelchair control – installation and programming
Menu tree
Input Devices
> Handcontrol Settings
> Drive
> Profile 1
> Latch Reverse
Input Devices
> Handcontrol Settings
> Drive
> Profile 1
> Power
64 | Ottobock
Access Min. Max.
De-
fault
Service
0
2
Service
20
100
100
Unit
0
Constant speed forward/reverse
If "Constant speed" = cruise is
selected, the wheelchair accelerates to
the desired speed. If the command is
reduced, the wheelchair maintains the
speed. Direction changes are always
possible. The wheelchair decelerates
if an opposing command of 50% or
less is issued. A stop is achieved via a
mode button command or an opposing
command of more than 50%.
If "Constant speed" = stepped is
selected, the wheelchair accelerates
by one increment with each short
command (33%, 66% and 100%)
up to the maximum speed. Direction
changes are always possible. The
wheelchair decelerates by one incre-
ment (e.g. from 66% to 33%) when a
short opposing command of 50% is
issued. A stop is achieved via a mode
button or with a permanent opposing
command.
Risk of injury due to incor-
WARNING
rectly installed control buttons. Never
select "Constant speed" mode (cruise
or stepped) unless the mode button is
installed so that the user can operate it
quickly at any time.
%
Power
The power parameter may be less
than 100% to limit the maximum
torque.
Limiting the torque reduces the power
of the wheelchair. This can prevent
dangerous collisions or crossing
obstacles.
Reducing torque is usually only feasi-
ble for indoor operation.
After reducing the torque, ensure the
user is able to cross any obstacles
that need to be crossed.
Help text
Skippi Plus

Advertisement

Table of Contents
loading