Auto Standby Timer; Unit Under Counter (Factory Set To "Off"); Display Contrast; Set Units - Thermo Scientific 1403 Series Operating Manual

Barnstead tii water purification system series 1403 and 1404
Table of Contents

Advertisement

Using your TII
Setting System Options
Options

Printer Attached

Reservoir Type

Auto Standby Timer

The AUTO STANDBY Timer can be set so that the unit will automatically be placed in STANDBY
and come back out of STANDBY and be placed into RUN mode at a specific time each day.

Unit Under Counter (Factory set to "OFF")

This option will lock out any type of automatic dispense out of the unit (Accudispense accessory
D13661 will still operate normally.) This will prevent accidental dispensing from the unit if it is
located under a counter or similar area such that the unit dispenser is not in site of the user but the
display is (i.e. remote mounted display.)

Display Contrast

The Thermo Scientific TII display can be adjusted based on user preference.

Set Units

The Thermo Scientific TII can display readings in a variety of units. Use this option to set desired
units.
Printer Attached
The Thermo Scientific TII can print readings to an optional printer accessory (AY1137X1).
Reservoir Type
The Thermo Scientific TII has been designed to interface with the Thermo Scientific TII storage
reservoir, as well as other reservoirs. If a reservoir other than a Thermo Scientific TII storage reservoir
will be used, the Reservoir Type must be set to "Alternate." This will configure the Thermo Scientific
TII system to communicate with two floats, 1 medium and 1 high. The float configuration must be
configured with the bottom float (medium) NORMALLY CLOSED – makes contact when there is
no water in the tank – and the top float (high) NORMALLY OPEN – does not make contact when
there is no water in the tank.
• Bottom float closed and top float open – "LOW"
• Bottom float open and top float open – "MED"
4-8
TII Water System
Procedure
MENU → SYSTEM OPTIONS → PRINTER ATTACHED → Choose
YES or NO → OK
MENU → SYSTEM OPTIONS → RESERVOIR TYPE → Choose DIA-
MOND TII or ALTERNATE → OK
Note When "UNIT UNDER COUNTER" is set to "ON," Auto Dispense will not be
available from the main unit dispenser, though the manual dispense knob is still
operational.