Diagnostics - GE akta start Maintenance Manual

Hide thumbs Also See for akta start:
Table of Contents

Advertisement

5 Service and settings
5.5 Pump

5.5.1 Diagnostics

5.5.1
Diagnostics
Parameter description
Parameter
Flow rate
Pump run
Check the Pump flow rate
Follow the instructions below to perform Pump diagnostics.
Step
1
2
3
76
Description
Setting of the desired flow rate (ml/min).
Displays the actual number of run hours for the Pump.
Action
Immerse the buffer inlet tubing A in demineralized water.
In the Pump screen, tap Diagnostics.
Result: The Diagnostics screen opens.
Use the up/down arrows to set the desired Flow Rate value in the range 0.5
to 5 ml/min
ÄKTA start Maintenance Manual 29-0603-08 AC