ÄKTA start Maintenance Manual 29-0603-08 AC
Action
Note the current temperature of the calibration solution in the Conductivity
flow cell as displayed in the Cell Temperature field.
In the Calibration screen, enter the conductivity value at the current tem-
perature in the Set Conductivity field and then tap Calibrate to carry out
the Conductivity calibration.
Note:
•
If a certified conductivity standard solution is used, use the supplied theo-
retical conductivity value corresponding to the temperature in question.
•
If a manually prepared 1.00 M NaCl calibration solution is used, get the
conductivity value at the current temperature from the graph for conduc-
tivity of the 1.00 M NaCl as a function of temperature presented below.
5 Service and settings
5.10 Conductivity Monitor
5.10.3 Conductivity flow cell calibration
99