Power failure
The result of a power failure depends on which unit is affected.
Power failure to...
ÄKTA start
UNICORN start on a
computer
ÄKTA start Maintenance Manual 29-0603-08 AC
will result in...
The run is interrupted immediately.
The data collected up to the time of the power fail-
ure is available on the USB memory stick or, if the
system is connected to a computer, UNICORN start.
The computer with UNICORN start installed shuts
down.
On the ÄKTA start Instrument Display, all four touch
buttons will be highlighted.
The run is interrupted immediately.
Data generated up to 10 seconds before the power
failure can be recovered.
Note:
The UNICORN start client may close down during a
temporary overload of the processor. This may appear
as a computer failure. The run continues and you can
restart the UNICORN start client to regain control.
2 Safety instructions
2.3 Emergency procedures
29