5.14.2

Export system report to USB

Instructions
Follow the instructions below to export the system report to a USB memory stick. Use
the system report in further contacts with GE Service Engineers.
Step
1
2
3
4
System report files parameters
The content of the INSTLOG.TXT file contains the following parameters:
The content of the ERRORLOG.TXT file contains Error codes, for example:
ÄKTA start Maintenance Manual 29-0603-08 AC
Action
Plug the USB memory stick into the USB port located on ÄKTA start.
In the System screen, tap Export System Report to USB.
Result: The files INSTLOG.TXT and ERRORLOG.TXT are exported to the USB
memory stick.
Remove the USB memory stick from the USB port and connect it to a com-
puter.
Check the content of the system report files INSTLOG.TXT and ERRORLOG.TXT,
see section System report file parameters.
ÄKTA start serial no.
Firmware version
FPGA version
Pump run
Pump tube run
UV LED run time
Buffer valve count
Wash valve count
Outlet valve count
Sample valve count
501: Over Pressure
301: Fraction collector failure
603: Illegal operation, restart instrument