Step
7
8
9
ÄKTA start Maintenance Manual 29-0603-08 AC
Action
Stop the Pump:
In the Settings and service screen, access Pump:Diagnostics screen.
•
Stop the Pump by tapping Stop flow.
•
If these checks indicate that the valve is faulty, replace the Sample valve.
If the valve has been replaced, tap Reset to set the Valve switches counter
to 0.
Result: A confirmation screen opens.
Tap Yes to confirm the reset if the valve has been replaced.
Or,
If the valve has not been replaced, tap No to cancel the action.
5 Service and settings
5.7 Sample valve
87