Pump - GE akta start Maintenance Manual

Hide thumbs Also See for akta start:
Table of Contents

Advertisement

6 Troubleshooting

6.4 Pump

6.4

Pump

Error messages
Error code
401
-
134
Descrip-
Possible cause
tion
Pump fail-
Pump is not working.
ure
No flow
The rollers are not rotat-
from the
ing.
Pump.
Action
Contact a GE Service
Engineer.
See
Troubleshooting Flow
chart 5, on page 143
Check the condition of
the pump tubing.
Check that there is no
blockage on the inlet
(or outlet).
Open the pump head
and visually check if
the pump rollers are
rotating freely when
starting the Pump
from the Diagnostics
screen.
If not, check if the
Pump connector is
connected or if any-
thing is stuck between
the rollers.
ÄKTA start Maintenance Manual 29-0603-08 AC

Hide quick links:

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the akta start and is the answer not in the manual?

Questions and answers