Step
7
8
9
ÄKTA start Maintenance Manual 29-0603-08 AC
Action
Replace with a new Conductivity Monitor and reconnect the tubing to the
ports.
Switch on the instrument by pressing the Power Switch to the l position.
In the Settings and service screen, access the Conductivity screen to set
the cell constant for the new Conductivity flow cell. For details, see
Service and settings, on page
55.
7 Disassembly
7.10 Conductivity Monitor
Chapter 5
185