6 Troubleshooting
6.1 UV
Warning
code
114
115
116
Error messages
Error code
101
130
Descrip-
Possible cause
tion
UV
In the Settings and ser-
reached
vice:UV screen:
end of life-
When trying to save the
time
UV light strength after
calibration, if the UV light
strength is 1020.
Flush UV
In the Settings and ser-
flow cell
vice:UV screen:
and mount
When trying to calibrate,
securely
if there are repeated cali-
brations and the signal
strength is decreasing.
UV base-
In the Settings and ser-
line ig-
vice:UV screen:
nored
Calibration and run time:
When the UV signal is not
in the range of 2500 mV
and 4000 mV, auto zero
cannot be performed.
Descrip-
Possible cause
tion
UV module
During power up, too little
failure
light is reaching the detec-
tor, i.e., too high ab-
sorbance in the cell or to
weak light source.
Action
Order a new UV Moni-
tor.
Perform UV LED cali-
bration.
For instructions, see:
Section 5.4.1 UV LED cali-
bration, on page 64
Section 5.4.3 Flow cell path
length, on page 69
Flush the UV flow cell
and mount it in a se-
cure way.
Recalibrate.
Flush the UV flow cell
and mount it in a se-
cure way.
Try to recalibrate. If
the voltage is still less
than 2500 mV, change
UV module.
Action
See
Troubleshooting Flow
chart 1, on page 139
ÄKTA start Maintenance Manual 29-0603-08 AC