6.5
Pressure sensor
Error messages
Error code
501
502
ÄKTA start Maintenance Manual 29-0603-08 AC
Descrip-
Possible cause
tion
Overpres-
Blockage in the tubing,
•
sure
valves or in the col-
umn.
Improperly selected
•
Flow rate.
Wrong pressure set-
•
ting used for chosen
application.
Pressure
The sensor is not connect-
sensor fail-
ed or not calibrated.
ure
6 Troubleshooting
6.5 Pressure sensor
Action
Check tubing and
•
valves by disconnect-
ing one at a time
starting from the
Fraction collector go-
ing backwards to-
wards the Pump.
When pressure is re-
leased the blockage
has been found, clean
or replace tube to re-
solve the issue.
Clean the column/sep-
•
aration media with
suitable solution (1 M
NaOH) or replace the
column with a new
one.
Check the specifica-
•
tion and pressure of
the columns for cor-
rect flow rate.
Perform Zero offset.
•
Contact a GE Service
•
Engineer.
135