Pressure Sensor - GE akta start Maintenance Manual

Hide thumbs Also See for akta start:
Table of Contents

Advertisement

6.5

Pressure sensor

Error messages
Error code
501
502
ÄKTA start Maintenance Manual 29-0603-08 AC
Descrip-
Possible cause
tion
Overpres-
Blockage in the tubing,
sure
valves or in the col-
umn.
Improperly selected
Flow rate.
Wrong pressure set-
ting used for chosen
application.
Pressure
The sensor is not connect-
sensor fail-
ed or not calibrated.
ure
6 Troubleshooting

6.5 Pressure sensor

Action
Check tubing and
valves by disconnect-
ing one at a time
starting from the
Fraction collector go-
ing backwards to-
wards the Pump.
When pressure is re-
leased the blockage
has been found, clean
or replace tube to re-
solve the issue.
Clean the column/sep-
aration media with
suitable solution (1 M
NaOH) or replace the
column with a new
one.
Check the specifica-
tion and pressure of
the columns for cor-
rect flow rate.
Perform Zero offset.
Contact a GE Service
Engineer.
135