GE akta start Maintenance Manual page 68

Hide thumbs Also See for akta start:
Table of Contents

Advertisement

5 Service and settings
5.4 UV Monitor
5.4.2 Diagnostics
Step
2
3
68
Action
Tap Stray light test.
The result is displayed as Pass or Fail:
Pass: the UV Monitor is working as it should.
Fail: clean, reassemble and secure the UV flow cell and the connected
tubing. Make sure that the detector is not exposed to e.g., direct sunlight.
If the test fails again, replace the flow cell and then possibly the UV
Monitor, or contact a GE Service Engineer.
Flush the system with demineralized water so that the Abs value returns
close to zero.
ÄKTA start Maintenance Manual 29-0603-08 AC