Step
3
4
5
Check for presence of stray light
Follow the instructions below to perform a Stray light test.
Step
1
ÄKTA start Maintenance Manual 29-0603-08 AC
Action
Tap Diagnostics.
Result: The Diagnostics screen opens.
Tap Set baseline to capture a new reference value for the test.
Tap Dark current test.
The result is displayed as Pass or Fail:
Pass: the UV Monitor is working as it should.
Fail: replace the UV Monitor.
Action
Flush and fill the UV flow cell with 15% acetone using a syringe. Make sure
that there are no air bubbles in the syringe.
5 Service and settings
5.4 UV Monitor
5.4.2 Diagnostics
67