7.8

Wash valve

Required tools
Tool
Torx driver
Instructions
Follow the instructions below to remove and replace the Wash valve.
Step
1
2
ÄKTA start Maintenance Manual 29-0603-08 AC
WARNING
Disconnect power. Always switch off power to the instrument be-
fore replacing any component on the instrument or cleaning the
instrument, unless stated otherwise in the user documentation.
Action
Switch off the instrument by pressing the Power Switch to the O position.
Remove the inlet and outlet tubing from the ports.
Dimension
T10
7 Disassembly

7.8 Wash valve

175