5.12.2
Diagnostics
Parameter description
Parameter
Feed tube test
Home test
Instructions
Follow the instructions below to run a diagnostics test on Frac30.
Step
1
2
ÄKTA start Maintenance Manual 29-0603-08 AC
Description
Checks that Frac30 rotates the Bowl assembly correctly
and shifts one tube at a time.
Checks that Frac30 rotates the Bowl assembly correctly
and shifts from the current position to the home position
(tube number 1).
Action
In the Fraction collector screen, tap Diagnostics.
Result: The Diagnostics screen opens.
Tap Feed tube test and visually observe that Frac30 rotates from one tube
to the next.
5 Service and settings
5.12 Frac30
5.12.2 Diagnostics
105