Mixer - GE akta start Maintenance Manual

Hide thumbs Also See for akta start:
Table of Contents

Advertisement

7.2

Mixer

Required tools
Tool
Torx screwdriver
Instructions
Follow the instructions below to remove and replace the Mixer.
Step
1
2
ÄKTA start Maintenance Manual 29-0603-08 AC
WARNING
Disconnect power. Always switch off power to the instrument be-
fore replacing any component on the instrument or cleaning the
instrument, unless stated otherwise in the user documentation.
Action
Switch off the instrument by pressing the Power Switch to the O position.
Remove the inlet and outlet tubing from the ports.
Dimension
T20
7 Disassembly

7.2 Mixer

153

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents