7.1 Buffer valve
Step
3
4
150
Action
Loosen the M3 screw from the valve using a T10 torx driver, supplied with
the equipment at delivery.
Note:
The screw can be located on different places on the top circumference of the
holder.
Slowly remove the Buffer valve until a cable is visible. The cable is assembled
with two connectors interconnected.