5 Service and settings
5.10 Conductivity Monitor
5.10.3 Conductivity flow cell calibration
5.10.3
Conductivity flow cell calibration
Prerequisites
Calibration solution:
•
•
Instructions
Follow the instructions below to calibrate the Conductivity flow cell.
Note:
Step
1
2
98
1.00 M NaCl
or,
100 mS/cm conductivity standard solution
Temperature compensation is enabled by default (factory setting). If you want
to disable temperature compensation, tap the Enable temperature compen-
sation checkbox in the Calibration screen.
Action
Fill the Conductivity flow cell with conductivity standard solution.
In the Conductivity screen, tap Calibration.
Result: The Calibration screen opens.
Note:
Make sure that the checkbox for Enable Temperature Compensation is
checked.
ÄKTA start Maintenance Manual 29-0603-08 AC