5 Service and settings
5.9 Outlet valve
Step
3
4
5
6
92
Action
In the Settings and service screen, tap Next to access the 2
Result: The following screen opens.
In the Settings and service screen, tap Outlet valve.
Result: The Outlet valve screen opens.
Check that the valve position Waste is selected.
Tap Turn valve to check that the Outlet valve switches between Waste
and Collection.
There is also a click sound when the valve switches.
Check that the flow is diverted to the Waste outlet or the Fraction col-
lector outlet according to the selected valve position.
Note:
The Wash valve is by default in Waste valve position. Hence, the flow is di-
verted to Waste.
Visually inspect the flow at the outlet. If the water is not flowing properly,
check the following and fix accordingly:
Is there a blockage in the tubing?
Are the connectors tightly connected?
Stop the Pump:
In the Settings and service screen, access Pump:Diagnostics screen.
Stop the Pump by tapping Stop flow.
nd
ÄKTA start Maintenance Manual 29-0603-08 AC
screen.