Chapter 17 Troubleshooting

17.2.2 Display and solutions for MainteTool

No.
Display Message
1
This application has already been
activated.
2
Current PATH get error
3
MainteTool.ini file does not exist.
Please create the file.
4
IP address of PC for PLNK5 is not
set.
5
Socket Initialization Error.
This application will be exit.
6
Socket Connection Error.
This application will be exit.
7
Please input a password.
8
Password is invalid.
9
Login Failure.
10
Login Failure (Password
mismatch for DRYPRO832-1).
11
Login Failure (Other user logged
in DRYPRO832-1).
12
Login Failure (Password
mismatch for DRYPRO832-2).
13
Login Failure (Other user logged
in DRYPRO832-2).
14
Login Failure (PLNK5).
15
Login Failure (DRYPR832-PC).
16
Login Failure (DRYPRO832-1).
17
Login Failure (DRYPRO832-2).
18
Login Failure (Other user logged
in PLNK).
19
Login Failure (Other user logged
in DRYPRO832-1).
Login Failure (Other user logged
in DRYPRO832-2).
20
Login Failure (Other user logged
in DRYPRO832-1).
Login Failure (DRYPRO832-2).
21
Login Failure (DRYPRO832-1).
Login Failure (DRYPRO832-2).
22
Login Failure (DRYPRO832-1).
Login Failure (Other user logged
in DRYPRO832-2).
17-58
Meaning
Double activation error
(tried to activate MainteTool, even though it was
already activated)
Current PATH get error
(failure to get current PATH at the startup)
No MainteTool.ini file (initial setup file) error
IP address of PLNK5-PC not set error
Socket initialization error
Socket connection error
No passowrd entered error
Password error
Login process failed
Login process failed
(password for DRYPRO832-1 does not match)
Login process failed
(other user already logged in to DRYPRO832-1)
Login process failed
(password for DRYPRO832-2 does not match)
Login process failed
(other user already logged in to DRYPRO832-2)
Login process failed
(failed on start request for PLNK5 maintenance)
Login process failed
(failed on start request for DRYPRO832PC
maintenance)
Login process failed
(failed on start request for DRYPRO832-1
maintenance)
Login process failed
(failed on start request for DRYPRO832-2
maintenance)
Login process failed
(other user already logged in to PLINK5)
Login process failed
(other user already logged in to DRYPRO832-1,
DRYPRO832-2)
Login process failed
(other user already logged in to DRYPRO832-1,
failed on start request for DRYPRO832-2
maintenance)
Login process failed
(failed on start request for DRYPRO832-1,
DRYPRO832-2)
Login process failed
(failed on start request for DRYPRO832-1, other
user already logged in to DRYPRO832-2)
Solution
Use the MainteTool that was activated
already.
Restart the PC.
UPGRADE to most current version.
UPGRADE to most current version.
Restart the PC.
Restart the PC.
Enter the password.
Enter the correct password.
UPGRADE to most current version.
UPGRADE to most current version.
Try again after the logged in user logs
out.
UPGRADE to most current version.
Try again after the logged in user logs
out.
UPGRADE to most current version.
UPGRADE to most current version.
UPGRADE to most current version.
UPGRADE to most current version.
Try again after the logged in user logs
out.
Try again after the logged in user logs
out.
Try again after other user logs out
from DRYPRO832-1
UPGRADE to most current version.
UPGRADE to most current version.
Try again after other user logs out
from DRYPRO832-2.
UPGRADE to most current version.