Replacing Position Regulation Inlet Sensor

Replace the photo sensor that detects film transportation at the inlet of the justification unit.
Be sure to turn off power and remove the power cable from the electrical outlet before removing
or installing the exterior cover or replacing loading parts.
Caution
Work outline
The number of personnel and the estimated work hours required to replace loading parts is as follows:
Number of
Personnel
1
Requirements
The procedure for replacing loading parts requires the following tools:
No.
1
Various types of screwdrivers
3
Photo sensor (for replacement)
Work Hours
15 minutes
Tools
1
2
Chapter 7 Replacing Sub Scan Unit Loading Parts
No.
2
Tray (container for screws and parts removed)
Remove the sub scan unit with reference to
"
Removal Procedure (Page 7-4)"
Removing/Installing Sub Scan Unit (Including the
Exposure Unit and Position Regulation
When replacing loading parts, be sure to
replace with the tray shutter attached. If
Caution
some of the film remains on the tray, the
film may be exposed.
When pulling out the sub scan unit, be
careful not to get it hung up on a wire.
The sub scan unit weighs approximately
16kg.
Carefully perform work when unloading
the sub scan unit from the main unit or
transporting it to work site.
If you place the removed sub scan unit on a
table, it may become extremely unstable.
It should be kept in mind that applying
excessive pressure could overturn the unit.
Refer to
step 2
and
3
in
"8.2 Replacing Exposure
Unit (Page 8-3)"
and remove the exposure unit.
The exposure unit includes optical parts.
Be careful not to subject to a strong impact.
Caution
Tools
in
"7.2
Unit)".
7-15