Replacing Internal Temperature Sensor - Konica Minolta drypro 832 Service Manual

Laser imager
Hide thumbs Also See for drypro 832:
Table of Contents

Advertisement

Chapter 9 Replacing Loading Parts in Heat Process Unit
9.2.6

Replacing Internal Temperature Sensor

Replace the temperature sensor that monitors the temperature inside the supply tray.
Be sure to turn off power and remove the power cable from the electrical outlet before
removing or installing the exterior cover or replacing loading parts.
Caution
The heat process unit and the area around it are hot immediately after turning off DRYPRO
832. The upper part of the rack is particularly hot. Touching it with bare hands may cause
burns. Therefore, use working gloves, and hold the grips on the side panels.
Work outline
The number of personnel and the estimated work hours required to replace loading parts is as follows:
Number of
Personnel
1
Requirements
The procedure for replacing loading parts requires the following tools:
No.
1
Various types of screwdrivers
3
Internal temperature sensor (for
replacement)
5
Snap band
9-22
Work Hours
20 minutes
Tools
1
Remove the left cover and heat process unit with
reference to
in
Caution
2
Remove the rear cover.
• Six screws (M4×8)
Note
No.
Tools
2
Tray (container for screws and parts
removed)
4
Nipper
6
Cotton work gloves
"
Removal Procedure (Page 9-3)"
"9.1.1 Removing/Installing Heat Process
When replacing loading parts, be sure to
replace with the tray shutter attached. If
some of the film remains on the tray, the
film may be exposed.
The heat process unit weighs
approximately 15kg.
Carefully perform work when unloading
the heat process unit from the main unit
or transporting it to work site.
Be very careful, the top of the rack is hot.
Remove the upper two screws and the one screw
in the center from the rear cover and loosen the
bottom three screws.
Unit".