Chapter 17 Troubleshooting
17.3.3 Replacing the DRYPRO 832
Method for replacing the DRYPRO 832 is different depending on the software version of the DRYPRO
832.
Note
•
Refer to
V1.01.
•
Refer to
V1.02 or later.
Replacing DRYPRO 832 (Up to V1.01)
This procedure is the replacing method when the software version of the DRYPRO 832 is up to V1.01.
Refer to
Note
V1.02 or later.
For information about the software version of the DRYPRO 832, refer to
25)".
Follow the procedure below to replace the DRYPRO 832.
Do not use the compact flash from the malfunctioning DRYPRO 832 in the new DRYPRO 832
main unit.
Caution
Requirements
Prepare following when replacing DRYPRO 832.
No.
1
Replacement DRYPRO 832 (V1.01 or later)
3
Backup file (DRYPRO 832)
5
DRYPRO 832 Installation Manual
START
Unpacking and
Installation
Exiting DRYPRO 832
main unit
Starting the the server PC Start the server PC (Printlink5-IN, CS-2, or CS-3)
Exit the application
(Only for CS-2 and CS-3)
17-82
"
Replacing DRYPRO 832 (Up to V1.01) (Page 17-82)"
"
Replacing the DRYPRO 832 (V1.02 or Later) (Page 17-90)"
"
Replacing the DRYPRO 832 (V1.02 or Later) (Page 17-90)"
Requirements
Refer to
"Chapter 3 Unpacking and Installation" in the Installation
DRYPRO 832.
Next, add supply trays or change supply tray sizes as required.
Refer to
"3.2.1 How to Exit" in the operation manual
power source breaker OFF.
Connection error message is displayed when the server PC is CS-2 or CS-3.
Click the [Yes] button, and close the dialog box.
Caution
Refer to
"7.4 Closing CS-2 or CS-3" in the installation manual
if the software version is up to
if the software version is
if the software version is
"15.11 System Setup (Page 15-
No.
Requirements
2
Upgrading file (V1.01 or later)
4
DRYPRO 832 Operation Manual
6
DRYPRO 832 Repair Manual
manual, and install the new
to exit the DRYPRO 832 and switch the
to exit CS-2 or CS-3.