15.15 HPRO Setup

Check or change the constants for controlling the heat process unit.
Each constant for HPRO setup is set to an optimum value at the factory. Do not change unless
otherwise instructed.
Caution
1
Refer to
"15.1 DRYPRO 832 Utility Tool Start (Page
window of the DRYPRO 832 Utility Tool.
2
Click the [HPRO SETUP] button under "SERVICE INFO".
The "HPRO SETUP" window appears.
"HPRO SETUP" window - "HPRO" tab
The "HPRO" tab in the "HPRO SETUP" window contains the following items.
No.
Setup Items
1
OFFSET (V1-V4)
2
HEAT BASE (V2,
V4, V5 QC)
Description of Settings
The offset value for each item of the heat process temperature control is displayed.
Display range: -250 - 250
The temperature measured from each section of the heat process guide is displayed.
Chapter 15 DRYPRO 832 Utility Tool
15-2)", and open the maintenance menu
15-33