Does Not Exit (Shutdown) - Konica Minolta drypro 832 Service Manual

Laser imager
Hide thumbs Also See for drypro 832:
Table of Contents

Advertisement

Phenomenon
Cannot connect on
the network

17.1.6 Does not exit (shutdown)

Phenomenon
The operation switch
was pressed and the
status LED has gone
out, but the DRYPRO
does not shut down
Parts to check
Use the ping
LAN cable disconnected
command from a
Depending on the type of
maintenance PC to
switching hub, the board may be
check the network
remembering the previous MAC
status
connection, and may therefore
not accept this product
Depending on the type of
switching hub, the connection
may not be established with auto-
negotiation
LAN board LED
The main board LAN function
does not flash
has a fault
A message "PC not
Communication fault between
connected" is
DRYPRO and server PC
displayed on the
DRYPRO 832
Parts to check
Communication
Main board fault or MEC
error between the
board fault
main board and
Main communication Assy
MEC board
connector disconnected or
broken wire
Communication
MEC board fault or multiple
error between MEC
power supply fault
board and multiple
Short circuit in remote signal
power supply
Assy
Chapter 17 Troubleshooting
Cause
Check that the LAN cable is connected.
Switch off the switching hub power, then
restart the hub.
The problem may not be fixed by only
disconnecting and reconnecting the LAN
cable.
Set the board settings for the switching
hub that connects this product to
100Base/TX, FDX or 1000Base/T.
Remove the installation location base,
and if a connection cannot be established
between a hand-held HUB and the PC,
replace the main board.
Check that power supply of the server
PC is ON.
Check the connection of the LAN
cable.
Restart the DRYPRO 832 and the
server PC.
Cause
Press and hold the operation switch once
again for more than 10 seconds.
If the DRYPRO does shut down, check
the connection of the main
communication Assy, and if there are no
problems, replace the main board, MEC
board, and main communication Assy.
If the DRYPRO does not shut down, there
is a communication error between the
MEC board and multiple power supply.
Switch the main power supply OFF, and
follow the "Communication error
between MEC board and multiple power
supply" solution below. (When a forced
shutdown can be performed)
Replace the MEC board, multiple power
supply, and remote signal Assy.
(When a forced shutdown cannot be
performed)
Solution
Solution
17-5