Konica Minolta drypro 832 Service Manual page 157

Laser imager
Hide thumbs Also See for drypro 832:
Table of Contents

Advertisement

7
Install the bracket removed in
• Two screws (M3×6)
Fix so that the positioning hole in the bracket aligns with the emboss on the mating part.
8
Connect the connectors (JP58) and (JP35) removed in
Be sure to push the connector all the way in when connecting it.
If the connector is not connected properly, an error will occur and, in some cases, smoke and/or
Caution
an odor may be generated.
9
Refer to
step 6
and
7
in
10
Install the sub scan unit with reference to
Installing Sub Scan Unit (Including the Exposure Unit and Position Regulation
11
Return the supply tray and tray shutter to their original positions.
Take sufficient care when handling the tray shutter, as it may be impossible to insert if it
becomes deformed due to rough handling.
Caution
This is the end of the process of replacing the Sub Scan Nip HP sensor.
Chapter 7 Replacing Sub Scan Unit Loading Parts
3
Place the sub scan unit upside down and remove
the connectors (JP58) and (JP35).
Caution
4
Remove the bracket.
• Two screws (M3×6)
5
Push the tab on the back of the sensor inward
and remove the photo sensor.
6
Insert a new photo sensor in the mounting hole.
step
4.
"8.2 Replacing Exposure Unit (Page 8-3)"
"
Installation Procedure (Page 7-7)"
When removing the connector, do not pull by
the wires, but hold the tab on the connector
with your fingers.
Pulling by the wires too tightly can cause a
break.
step
3.
and install the exposure unit.
in
"7.2 Removing/
Unit)".
7-39