Chapter 1 Safety Precautions and Warnings
1.2

Safety Precautions

Read all safety precautions thoroughly before using the product.
Be sure to observe the safety precautions described in this section.

1.2.1 Precautions Following Requirements of Laws and Regulations

DRYPRO 832 Laser Unit Specifications
Item
Wavelength
Maximum output
(1) EMC Statement - DRYPRO 832 has been tested and found to comply with the limits for medical devices to the
IEC 60601-1-2: 2001, Medical Device Directive 93/42/EEC or the Electromagnetic Compatibility Directive 89/
336/EEC (use applicable directive). These limits are designed to provide reasonable protection against harmful
interference in a typical medical installation. DRYPRO 832 generates, uses and can radiate radio frequency
energy and, if not installed and used in accordance with the instructions, may cause harmful interference to other
devices in the vicinity. However, there is no guarantee that interference will not occur in a particular installation.
If DRYPRO 832 does cause harmful interference to other devices, which can be determined by turning the
equipment off and on, the user is encouraged to try to correct the interference by one or more of the following
measures:
- Reorient or relocate the receiving device.
- Increase the separation between the equipment.
- Connect the equipment into an outlet on a circuit different from that to which the other device(s) are connected.
- Consult the manufacturer or field service technician for help.
(2) EMC Statement - Take precautions against DRYPRO MODEL 832 especially regarding EMC. Install and put
into service according to the EMC information provided in the manual.
(3) EMC Statement - Do not use mobile phones or pocket pagers in the vicinity of the DRYPRO MODEL 832. Use
of mobile phones or pocket pagers near the DRYPRO MODEL 832 can cause errors in operation due to
electromagnetic wave interference : such devices should be turned off in the vicinity of the unit.
(4) EMC Statement - Cable List
- AC Power Cable (3m)
(5) EMC Statement - The use of accessories, transducers and cables other than those specified, with the exception of
transducers and cables sold by KONICA MINOLTA MEDICAL & GLAPHIC, INC. as for internal components,
may result in increased emissions or decreased immunity of DRYPRO 832.
(6) EMC Statement - Do not use adjacent to or stacked with other equipment. If adjacent or stacked use is necessary,
confirm normal operation in the configuration in which DRYPRO 832 will be used.
1-4
Caution as to laser control
To prevent danger, make sure that only service engineers who received proper formal
training will remove the exterior cover or touch the inside of the equipment.
784nm
260mW