16.6 Testing Voltage Transformer for RECARO Seat
Testing Preparations
Remove battery boxes. See chapter 12.1 in this connection.
Remove splash protection. See chapter 11.7 in this connection.
Install battery boxes.
Testing Voltage on Voltage Transformer for RECARO seat
Detach plug connection from the battery cable to the
voltage transformer.
Measure the voltage existing at the contacts of the feeder
cable (see drawing).
It must amount to at least 19 volts.
Reconnect the plugs.
Detach the plug connection to the RECARO seat.
Measure the output voltage.
Depending on the charging status of the batteries, it must
amount to between 2.5 and 4 volts.
NOTE:
The voltage transformer provides an output voltage of 12 volts only under load
(seat connected and switched on). In conditions without load, only a level of
between 2.5 and 4 volts is to be measured.
CAUTION:
Malfunction
No voltage at the
feeder cable?
Voltage below 19 V?
Output voltage
higher than 4 volts and
lower than 2.5 volts?
90
Avoid by all means bridging the contacts of the power supply.
>Danger of Fire and Burning<.
Feeder cable defective.
Fuse (15 A) defective.
Power supply to the main module is
interrupted.
Battery fully discharged.
Voltage transformer defective.
Cause
Check / replace feeder cable.
Replace fuse.
Check voltage supply.
Check master fuse (50 A).
Pre-charge battery.
Replace battery.
Replace voltage transformer.
Main Module and Voltage Transformer
Fuse 15A
Plug of
feeder cable
Feeder : Testing Voltage
+
min. 19 volts
Remedy
Plug for
RECARO seat
V
Chapter
18.0
12.1
12.1
12.2
12.0
17.7