17.6 Main Module

Preparation
• Remove battery boxes. See chapter 12.1 in this respect.
• Remove splash protection. See chapter 11.7 in this repsct.
• Remove cable binders located at the main module.
Removing the Main Module
• Remove all plugs from the socket board of the main module.

NOTE:
Press down plug lock, as otherwise the plugs will be
damaged.
• Loosen the two fastening screws of the main module
(SW 8 mm).
• Pull module out of the battery box retainer in downward direction.
Installing the Main Module:
• Push the main module into the battery box retainer from below und
and fasten it by means of the two screws.
• Insert the plugs into the module’s socket board .
The arrangement of the plugs can be seen from the connecting
plan.
• Attach new cable binders.
• Mount the splash protection.
• Insert the battery boxes.
Connecting Plan: Main Module
linker Fahrmotor
left driving motor
M 2
Batterie
ACS-Fahrpult
Battery
ACS-Remote
+
+
1
D
+
X
B
U
S
24V
zu weiteren Modulen
rechter Fahrmotor
right driving motor
M 1
2
Location of the cable binders
Removing the main module
Fastening screw
M5 x 20 - SW 8
Detail: Module Fastening
Washer
Nut M5 - SW 8,
Washer
self-locking
Screw
M5 x 20 - SW 8
Cable
binders
101