Invacare STORM Service Manual page 18

Hide thumbs Also See for STORM:
Table of Contents

Advertisement

An actuator reacting
irregularly or not at all
Motor or adjustment
module defective
No flashing code?
Actuator defective
Servo/lighting module
defective
Joystick box defective
Adjustment control
defective
Bbus cable defective
Status display is flashing,
after 4 sec. the driving mode
display is flashing too
Joystick was not in
neutral position when
switched on
Joystick box defective
Lighting system defective
Defective lamp
Defective cable / plug
Defective
servo/lighting module
Wrong software
Battery charging
not possible
Defective battery
Defective charger
Defective horn
18
Check flashing code
on joystick box
Replace actuator
Replace servo/lighting
module
Replace joystick box
Replace adjustment
control
Replace bus cable
Switch joystick box
off and on again
Replace joystick box
Replace lamp
Check / replace
cable / plug
Replace servo/lighting
module
Renew
programming
Replace battery
Replace charger
Replace joystick box
Chapter 8.0
Chapter 14.3
Chapter 17.7
Chapter 17.1 - 17.5
Chapter 12.2
Chapter 12.2
Chapter 11.1 and 11.2
Chapter 11.1 and 11.2
Chapter 12.3
Consult Manual on
programming device
or PCD software
Chapter 10.1
Chapter 10.3
Chapter 12.2

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents