Disassembly (complete)
• Release legrest lock (5) and lift legrest from the hinge bolts (3 and 4).
• Loosen nut (20) and pull insert tube (21), together with footplate tube (22) and footplate (23), from
legrest elbow tube (9).
• Loosen screw (25) with washers (26) and nut (27) completely and separate insert tube (21) from
footplate tube (22).
• If required, release locking screw (24) and nut (28) and detach from footplate tube (22).
• Loosen tapping screws (29) completely and detach footplate (23) from footplate tube (22).
• Loosen nut (20) completely and detach together with washer (19), screw (18) and clamping piece (17)
from legrest elbow tube (9).
• Loosen two screws (38) and separate calfplate holder (31) from calfplate (32).
• Loosen nut (37) and, together with washer (36), bushing (34), screw (35) and calfplate holder (31, sepa-
rate from calfplate joint (30).
• Loosen screws and remove calfplate joint (30) from legrest elbow tube (9).
• Lift up starlock cap (16), for example with a screw driver, to release the guide rod (13) from the legrest
upper part.
• Loosen screw (11) and nut (12) completely and detach legrest lelbow part (9) and bushing (10) from
legrest upper part (6).
• Lift up Starlock cap (16), for example with a screw driver, to free the release lever (14).
• Pull release lever (14) from the legrest elbow tube (9) and detach the backlocking unit (39) together with
washers (15) and guide rod (13) from the legrest elbow tube (9) .
• Release (12) nut and separate guide rod (13) from the backlocking unit (39).

NOTE:
Do not lose / misplace the individual parts of the backlocking unit
(release plate, release block and spring).
• If necessary, remove slide pieces (7 and 8) and cap (33).

NOTE:
The slide pieces are inserted. They can be lifted up with a suitable tool,
such as a screwdriver.