16.5 Testing Additional Power Supply of the 12 V Lighting Module (TÜV Version)
NOTE:
If required, test model first according to the description given
>in chapter 16.4.
Testing Preparations
Remove battery boxes. See chapter 12.1 in this respect.
Remove splash protection. See chapter 11.7 in this respect.
Install battery boxes.
Detach the plug connection from the battery cable to the
12 V lighting module.
Measure the voltage existing at the contacts of the feeder
cable (see drawing).
It must amount to at least 19 volts.
CAUTION:
Malfunction
No voltage at contacts?
Voltage below 19 V?
Avoid by all means a bridging of the power supply contacts.
>Danger of Fire and Burning<.
Power supply to the plug of the main
module interrupted.
Battery fully discharged.
Cause
Check / replace cable.
Check power supply.
Test master fuse (50 A).
Pre-charge battery.
Replace battery.
Main Module and 12 V Lighting Module
Plug vonnection
Additional Cable: Testing Voltage
+
Remedy
V
min. 19 volts
Chapter
18.0
12.0
12.0
12.2
12.0
89