Removing Battery Shroud
1.
Press safety lock on top of battery box A and remove
handle.
2.
Remove battery shroud B from battery box.
Installing Battery Shroud
1. Install battery shroud.
2. Install handle to shroud. Make sure safety lock engages.

8.8 Electrical components

8.8.1 Replacing Power Module

The various power modules can be installed to the mobility
device with a range of different remotes. Power modules
are described in corresponding controls manuals or earlier
revisions of this manual. Replacement is described below
using LiNX power module as an example. The course of
action is the same for other power modules. The only
difference is the terminal layout.
CAUTION!
Any changes to the drive program can affect the
driving characteristics and the tipping stability
of the mobility device
– Changes to the drive program may only be
carried out by trained providers.
– Invacare can only give a warranty for safe
mobility device driving behaviour - especially
the tipping stability - for unaltered standard
drive programs.
All power modules are delivered with a standard drive
program. If you have carried out customer-specific
modifications to the drive program, you will have to
make these changes again after installing new power
module.
1577880-C
Removing Power Module
Phillips screwdriver, size 2
To adapt the drive program: Programming
software or hand programming device and
system installation manual, available from
Invacare.
1. Remove rear shroud. See 8.7.1 Rear Shroud, page 20.
2. Remove all plugs A from power module.
Mark positions of individual plugs for later
re-installing.
3. Loosen and remove screws B on both sides of power
module.
4. Remove power module.
Installing Power Module
For details on cable routing and information about
Updating drive program and Adapting drive program,
see LiNX Service Manual.
1. Install parts in reverse order.
2. If LiNX system is used, perform calibration process.
See Suggested programming procedure in LiNX service
manual.
3. If available, update drive program.
4. If necessary, adapt drive program to user's need.
5. Test all functions.

8.8.2 Batteries

CAUTION!
Injury hazard and possible material damages if
batteries are handled improperly
The installation of new batteries may only be
carried out by authorised specialists.
– Observe the warning information on the
batteries.
– Only use battery versions stated in the
specifications.
CAUTION!
Fire and burns hazard if battery terminal is
bypassed
– Take great care to ensure that the battery
terminals are never short-circuited with tools or
mechanical mobility device parts.
– Ensure that the battery terminal caps have
been replaced if you are not working on the
battery terminals.
Fig. 8-2
Service
21