Remote Cables - Agilent Technologies InfinityLab LC Series User Manual

Isocratic and quaternary pumps
Hide thumbs Also See for InfinityLab LC Series:
Table of Contents

Advertisement

10
Identifying Cables

Remote Cables

Remote Cables

ERI (Enhanced Remote Interface)
p/n 5188-8029
Configuration is different with old firmware revisions.
NOTE
The configuration for IO4 and IO5 is swapped for modules with firmware lower
than D.07.10.
Peak Detection is used for LCMS systems connected with the Fraction Collection
NOTE
Remote Y-Cable (5188-8057).
Agilent InfinityLab LC Series 1260 Infinity II Isocratic and Quaternary Pumps User Manual
5188-8029 ERI to general purpose (D-Sub 15 pin male - open end)
5188-8044 ERI to ERI (D_Sub 15 pin male - male)
5188-8059 ERI-Extension-Cable 1.2 m (D-Sub15 pin male / female)
pin
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
NC
Color code
Enhanced
Remote
white
IO1
brown
IO2
green
IO3
yellow
IO4
grey
IO5
pink
IO6
blue
IO7
red
IO8
black
1wire DATA
violet
DGND
grey-pink
+5V ERI out
red-blue
PGND
white-green
PGND
brown-green
+24V ERI out
white-yellow
+24V ERI out
yellow-brown
Classic
Active (TTL)
Remote
START
Low
REQUEST
STOP
Low
READY
High
PEAK DETECT
Low
POWER ON
High
SHUT DOWN
Low
START
Low
PREPARE
Low