Replace The O-Ring On The Purge Valve - Agilent Technologies InfinityLab LC Series User Manual

Isocratic and quaternary pumps
Hide thumbs Also See for InfinityLab LC Series:
Table of Contents

Advertisement

8
Maintenance

Replace the O-Ring on the Purge Valve

Replace the O-Ring on the Purge Valve

When
If the original o-ring is damaged and needs to be replaced
Tools required
p/n
8710-0510
8710-1924
OR
Parts required
#
1
1
1
Preparations
1
Disassemble the purge valve.
Agilent InfinityLab LC Series 1260 Infinity II Isocratic and Quaternary Pumps User Manual
Description
A
Open-end wrench 1/4 — 5/16 inch
A
Open-end wrench 14 mm
Pair of tweezers
Toothpick
p/n
Description
A
5067-6595
1260 PV O-ring FKM 5/pack
A
01018-22707
PTFE frits (pack of 5) (OPTIONAL)
A
5067-4728
Seal cap (OPTIONAL)
Switch off pump at the main power switch.
Open the doors of the module.
Use an optional solvent shutoff valve or lift up solvent filters in solvent reservoirs for avoiding
leakages.
Remove the purge valve from the pump head.
2
Remove the old o-ring from the purge valve.
3
Clean the purge valve parts.
139