2
Site Requirements and Specifications
Performance Specifications
Table 7
Type
Composition precision
Integrated degassing unit
Instrument Control
Communication
Safety and maintenance
GLP features
Housing
Agilent InfinityLab LC Series 1260 Infinity II Isocratic and Quaternary Pumps User Manual
Performance Specifications G7111B
Specification
< 0.2 % RSD or < 0.04 min SD,
whichever is greater
Number of channels: 4
Internal volume per channel:
1.5 mL
LC & CE Drivers A.02.14 or above
Instrument Control Framework
(ICF) A.02.04 or above
InfinityLab LC Companion
(G7108AA) with firmware D.07.25
or above
Instant Pilot (G4208A) with
firmware B.02.20 or above
Lab Advisor software B.02.08 or
above
Controller Area Network (CAN)
Local Area Network (LAN)
Extended Remote Interface (ERI)
Universal Serial Bus (USB)
Extensive diagnostics, error
detection and display through
Agilent Lab Advisor, leak detection,
safe leak handling, leak output
signal for shutdown of the
pumping system.
Low voltage in major maintenance
areas.
Early maintenance feedback (EMF)
for continuous tracking of
instrument usage in terms of seal
wear and volume of pumped
mobile phase with pre-defined and
user settable limits and feedback
messages. Electronic records of
maintenance and errors
All materials are recyclable
Comments
At 1 mL/min; based on
retention time at constant
temperature
For details about supported
software versions refer to the
compatibility matrix of the LC &
CE Drivers version in use
44